Directive 2014/32/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of measuring instruments (recast) (Text with EEA relevance)

 Help about what version

Pa Fersiwn

 Help about advanced features

Nodweddion Uwch

 Help about opening options

Dewisiadau AgorExpand opening options

 Help about UK-EU Regulation

Legislation originating from the EU

When the UK left the EU, legislation.gov.uk published EU legislation that had been published by the EU up to IP completion day (31 December 2020 11.00 p.m.). On legislation.gov.uk, these items of legislation are kept up-to-date with any amendments made by the UK since then.

Changes over time for: ANNEX X

 Help about opening options

Status:

EU Directives are published on this site to aid cross referencing from UK legislation. Since IP completion day (31 December 2020 11.00 p.m.) no amendments have been applied to this version.

ANNEX XU.K.MATERIAL MEASURES (MI-008)

CHAPTER IU.K. Material measures of length

The relevant essential requirements of Annex I, the specific requirements of this Annex and the conformity assessment procedures listed in this chapter, apply to material measures of length defined below. However, the requirement for the supply of a copy of declarations of conformity may be interpreted as applying to a batch or consignment rather than each individual instrument.

DEFINITIONSU.K.

Material measure of lengthAn instrument comprising scale marks whose distances are given in legal units of length.

SPECIFIC REQUIREMENTSU.K.

Reference Conditions U.K.
1.1.For tapes of length equal to or greater than 5 metres, the maximum permissible errors (MPEs) are to be met when a tractive force of fifty newtons or other force values as specified by the manufacturer and marked on the tape accordingly, or in the case of rigid or semi-rigid measures no tractive force is needed, is applied.U.K.
1.2.The reference temperature is 20 °C unless otherwise specified by the manufacturer and marked on the measure accordingly.U.K.
MPEs U.K.
2.The MPE, positive or negative in mm, between two non-consecutive scale marks is (a + bL), where:U.K.
  • L is the value of the length rounded up to the next whole metre; and

  • a and b are given in Table 1 below.

When a terminal interval is bounded by a surface, the MPE for any distance beginning at this point is increased by the value c given in Table 1.

Table 1
a

Applies to the tape/dip weight combinations.

b

If the nominal tape length exceeds 30 m, an additional mpe of 0,75 mm shall be permitted for each 30 m of tape length.

Accuracy Classa (mm)bc (mm)
I0,10,10,1
II0,30,20,2
III0,60,40,3
D —special class for dipping tapesa

Up to and including 30 mb

1,5zerozero
S —special class for tank strapping tapes

For each 30 m length when the tape is supported on a flat surface

1,5zerozero

Dip tapes may also be of Classes I or II in which case for any length between two scale marks, one of which is on the sinker and the other on the tape, the MPE is ± 0,6 mm when application of the formula gives a value of less than 0,6 mm.

The MPE for the length between consecutive scale marks, and the maximum permissible difference between two consecutive intervals, are given in Table 2 below.

Table 2
Length i of the intervalMPE or difference in millimetres according to accuracy class
IIIIII
i ≤ 1 mm0,10,20,3
1 mm < i ≤ 1 cm0,20,40,6

Where a rule is of the folding type, the jointing shall be such as not to cause any errors, supplementary to those above, exceeding: 0,3 mm for Class II, and 0,5 mm for Class III.

Materials U.K.
3.1.Materials used for material measures shall be such that length variations due to temperature excursions up to ± 8 °C about the reference temperature do not exceed the MPE. This does not apply to Class S and Class D measures where the manufacturer intends that thermal expansion corrections shall be applied to observed readings where necessary.U.K.
3.2.Measures made from material whose dimensions may alter materially when subjected to a wide range of relative humidity, may only be included in Classes II or III.U.K.
Markings U.K.
4.The nominal value shall be marked on the measure. Millimetre scales shall be numbered every centimetre and measures with a scale interval greater than 2 cm shall have all scale marks numbered.U.K.

CONFORMITY ASSESSMENTU.K.

The conformity assessment procedures referred to in Article 17 that the manufacturer can choose between are:

F 1 or D1 or B + D or H or G.

CHAPTER IIU.K. Capacity serving measures

The relevant essential requirements of Annex I, and the specific requirements and the conformity assessment procedures listed in this chapter, apply to capacity serving measures defined below. However, the requirement for the supply of a copy of declarations of conformity may be interpreted as applying to a batch or consignment rather than each individual instrument. Also, the requirement for the instrument to bear information in respect of its accuracy shall not apply.

DEFINITIONSU.K.

Capacity serving measureA capacity measure (such as a drinking glass, jug or thimble measure) designed to determine a specified volume of a liquid (other than a pharmaceutical product) which is sold for immediate consumption.
Line measureA capacity serving measure marked with a line to indicate nominal capacity.
Brim measureA capacity serving measure for which the internal volume is equal to the nominal capacity.
Transfer measureA capacity serving measure from which it is intended that the liquid is decanted prior to consumption.
CapacityThe capacity is the internal volume for brim measures or internal volume to a filling mark for line measures.

SPECIFIC REQUIREMENTSU.K.

1. Reference Conditions U.K.
1.1.Temperature: the reference temperature for measurement of capacity is 20 °C.U.K.
1.2.Position for correct indication: free standing on a level surface.U.K.
2. MPEs U.K.
Table 1
LineBrim
Transfer measures
< 100 ml± 2 ml

– 0

+ 4 ml

≥ 100 ml± 3 %

– 0

+ 6 %

Serving measures
< 200 ml± 5 %

– 0

+ 10 %

≥ 200 ml± (5 ml + 2,5 %)

– 0

+ 10 ml + 5 %

3. Materials U.K.

Capacity serving measures shall be made of material which is sufficiently rigid and dimensionally stable to maintain capacity within the MPE.

4. Shape U.K.
4.1.Transfer measures shall be designed so that a change of contents equal to the MPE causes a change in level of at least 2 mm at the brim or filling mark.U.K.
4.2.Transfer measures shall be designed so that the complete discharge of the liquid being measured will not be impeded.U.K.
5. Marking U.K.
5.1.The nominal capacity declared shall be clearly and indelibly marked on the measure.U.K.
5.2.Capacity serving measures may also be marked with up to three clearly distinguishable capacities, none of which shall lead to confusion one to the other.U.K.
5.3.All filling marks shall be sufficiently clear and durable to ensure that MPEs are not exceeded in use.U.K.

CONFORMITY ASSESSMENTU.K.

The conformity assessment procedures referred to in Article 17 that the manufacturer can choose between are:

A2 or F1 or D1 or E1 or B + E or B + D or H.

Yn ôl i’r brig

Options/Help

You have chosen to open the Whole Directive

The Whole Directive you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?

You have chosen to open Schedules only

Y Rhestrau you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?