ANNEX XIU.K.

DIMENSIONAL MEASURING INSTRUMENTS (MI-009)U.K.

The relevant essential requirements of Annex I, the specific requirements of this Annex and the conformity assessment procedures listed in this Annex, apply to dimensional measuring instruments of the types defined below.

DEFINITIONSU.K.

Length measuring instrumentA length measuring instrument serves for the determination of the length of rope-type materials (e.g. textiles, bands, cables) during feed motion of the product to be measured.
Area Measuring InstrumentsAn area measuring instrument serves for the determination of the area of irregular shaped objects, e.g. for leather.
Multi-dimensional Measuring InstrumentsA multi-dimensional measuring instrument serves for the determination of the edge length (length, height, width) of the smallest enclosing rectangular parallelepiped of a product.

CHAPTER IU.K. Requirements common to all dimensional measuring instruments

Electromagnetic immunity U.K.

1.The effect of an electromagnetic disturbance on a dimensional measuring instrument shall be such that:U.K.
2.The critical change value is equal to one scale interval.U.K.
CONFORMITY ASSESSMENTU.K.

The conformity assessment procedures referred to in Article 17 that the manufacturer can choose between are:

CHAPTER IIU.K. Length measuring instruments

Characteristics of the product to be measured U.K.

1.Textiles are characterised by the characteristic factor K. This factor takes the stretchability and force per unit area of the product measured into account and is defined by the following formula:U.K.
K=

, where

ε is the relative elongation of a cloth specimen 1 m wide at a tensile force of 10 N,

GA is the weight force per unit area of a cloth specimen in N/m2.

Operating conditions U.K.

2.1. Range U.K.

Dimensions and K-factor, where applicable, within the range specified by the manufacturer for the instrument. The ranges of K-factor are given in Table 1:

Table 1
GroupRange of KProduct
I0 < K < 2 × 10–2 N/m2low stretchability
II2 × 10–2 N/m2 < K < 8 × 10–2 N/m2medium stretchability
III8 × 10–2 N/m2 < K < 24 × 10–2 N/m2high stretchability
IV24 × 10–2 N/m2 < Kvery high stretchability
2.2.Where the measured object is not transported by the measuring instrument, its speed must be within the range specified by the manufacturer for the instrument.U.K.
2.3.If the measurement result depends on the thickness, the surface condition and the kind of delivery (e.g. from a big roll or from a pile), corresponding limitations are specified by the manufacturer.U.K.

MPEs U.K.

3. Instrument U.K.
Table 2
Accuracy classMPE
I0,125 %, but not less than 0,005 Lm
II0,25 %, but not less than 0,01 Lm
III0,5 %, but not less than 0,02 Lm

Where Lm is the minimum measurable length, that is to say the smallest length specified by the manufacturer for which the instrument is intended to be used.

The true length value of the different types of materials shall be measured using suitable instruments (e.g. tapes of length). Thereby, the material which is going to be measured shall be laid out on a suitable underlay (e.g. a suitable table) straight and unstretched.

Other requirements U.K.

4.The instruments must ensure that the product is measured unstretched according to the intended stretchability for which the instrument is designed.U.K.

CHAPTER IIIU.K. Area measuring instruments

Operating conditions U.K.

1.1. Range U.K.

Dimensions within the range specified by the manufacturer for the instrument.

1.2. Condition of the product U.K.

The manufacturer shall specify the limitations of the instruments due to the speed, and thickness of the surface conditions if relevant, of the product.

MPEs U.K.

2. Instrument U.K.

The MPE is 1,0 %, but not less than 1 dm2.

Other requirements U.K.

3. Presentation of the product U.K.

In the case of pulling back or stopping the product, it shall not be possible to have an error of measurement or the display must be blanked.

4. Scale interval U.K.

The instruments must have a scale interval of 1,0 dm2. In addition, it must be possible to have a scale interval of 0,1 dm2 for testing purposes.

CHAPTER IVU.K. Multidimensional measuring instruments

Operating conditions U.K.

1.1. Range U.K.

Dimensions within the range specified by the manufacturer for the instrument.

1.2. Minimum dimension U.K.

The lower limit of the minimum dimension for all values of the scale interval is given in Table 1.

Table 1
Scale interval (d)Minimum dimension (min)(lower limit)
d ≤ 2 cm10 d
2 cm < d ≤ 10 cm20 d
10 cm < d50 d
1.3. Speed of the product U.K.

The speed must be within the range specified by the manufacturer for the instrument.

MPE U.K.

2. Instrument: U.K.

The MPE is ± 1,0 d.