http://www.legislation.gov.uk/uksi/2000/388/schedule/1The Weighing Equipment (Automatic Gravimetric Filling Instruments) Regulations 2000Weights and measuresWEIGHTS AND MEASURESStatute Law Database2024-04-02Expert Participation2020-12-31These Regulations implement, in Great Britain, International Recommendation OIML R 61 of the Organisation Internationale de Métrologie Légale relating to automatic gravimetric filling instruments (Edition 1996 (E)), to the extent that the Recommendation applies to such instruments following the grant or renewal of a certificate of approval of a pattern by the Secretary of State under section 12 of the Weights and Measures Act 1985 (the 1985 Act).SCHEDULE 1

regulation 5(a)

(Manner of erection and installation)Cleaning and testing

Every filling instrument shall be so positioned as to facilitate cleaning and testing.

Feeding device (Extract from Clause 3.5 of Part 1 of OIML R 61)

The feeding device shall be designed to provide sufficient and regular flowrate(s).

Load receptor (Clause 3.6 of Part 1 of OIML R 61)

The load receptor, and feed and discharge devices as appropriate, shall be designed to ensure that residual material retained after each discharge is negligible.

Instruments using the subtractive weighing principle shall be designed to ensure that residual material retained at feed from the discharge gate is negligible.

The load receptor shall provide access and facilities so that where necessary test weights or masses up to the maximum capacity can be placed in position, in a safe and secure manner. If these facilities are not a permanent fixture of the instrument, they must be kept in the vicinity of the instrument.

Manual discharge of the load receptor shall not be possible during automatic operation.

Equilibrium mechanism (Clause 3.8 of Part 1 of OIML R 61)

The equilibrium mechanism may be provided with detachable masses which shall be either weights in accordance with OIML requirements or purpose designed masses of any nominal value, distinguishable by shape and identified with the filling instrument.

Installation (Extract from Clause 5.3.1 of Part 1 of OIML R 61)

The installation of a filling instrument shall be so designed that an automatic weighing operation will be the same whether for the purposes of testing or for use for a transaction.

This XML file does not appear to have any style information associated with it. The document tree is shown below.
<akomaNtoso xmlns:uk="https://www.legislation.gov.uk/namespaces/UK-AKN" xmlns:ukl="http://www.legislation.gov.uk/namespaces/legislation" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns="http://docs.oasis-open.org/legaldocml/ns/akn/3.0" xsi:schemaLocation="http://docs.oasis-open.org/legaldocml/ns/akn/3.0 http://docs.oasis-open.org/legaldocml/akn-core/v1.0/cos01/part2-specs/schemas/akomantoso30.xsd">
<act name="uksi">
<meta>
<identification source="#">
<FRBRWork>
<FRBRthis value="http://www.legislation.gov.uk/id/uksi/2000/388"/>
<FRBRuri value="http://www.legislation.gov.uk/id/uksi/2000/388"/>
<FRBRdate date="2000-02-15" name="made"/>
<FRBRauthor href="http://www.legislation.gov.uk/id/government/uk"/>
<FRBRcountry value="GB-UKM"/>
<FRBRsubtype value="regulation"/>
<FRBRnumber value="388"/>
<FRBRname value="S.I. 2000/388"/>
<FRBRprescriptive value="true"/>
</FRBRWork>
<FRBRExpression>
<FRBRthis value="http://www.legislation.gov.uk/uksi/2000/388/2020-12-31"/>
<FRBRuri value="http://www.legislation.gov.uk/uksi/2000/388/2020-12-31"/>
<FRBRdate date="2020-12-31" name="validFrom"/>
<FRBRauthor href="#"/>
<FRBRlanguage language="eng"/>
</FRBRExpression>
<FRBRManifestation>
<FRBRthis value="http://www.legislation.gov.uk/uksi/2000/388/2020-12-31/data.akn"/>
<FRBRuri value="http://www.legislation.gov.uk/uksi/2000/388/2020-12-31/data.akn"/>
<FRBRdate date="2024-12-06Z" name="transform"/>
<FRBRauthor href="http://www.legislation.gov.uk"/>
<FRBRformat value="application/akn+xml"/>
</FRBRManifestation>
</identification>
<lifecycle source="#">
<eventRef refersTo="#made" date="2000-02-15" eId="date-made" source="#"/>
<eventRef refersTo="#laid" date="2000-02-17" eId="date-laid-1" source="#united-kingdom-parliament"/>
<eventRef refersTo="#coming-into-force" date="2000-07-17" eId="date-cif-1" source="#"/>
<eventRef date="2000-07-17" eId="date-2000-07-17" source="#"/>
<eventRef date="2018-09-01" eId="date-2018-09-01" source="#"/>
<eventRef date="2020-12-31" eId="date-2020-12-31" source="#"/>
</lifecycle>
<analysis source="#">
<restrictions source="#">
<restriction refersTo="#extent-e+w+s" type="jurisdiction"/>
<restriction href="#schedules" refersTo="#extent-e+w+s" type="jurisdiction"/>
<restriction href="#schedule-1" refersTo="#extent-e+w+s" type="jurisdiction"/>
<restriction href="#d24e80" refersTo="#extent-e+w+s" type="jurisdiction"/>
<restriction href="#d24e87" refersTo="#extent-e+w+s" type="jurisdiction"/>
<restriction href="#d24e97" refersTo="#extent-e+w+s" type="jurisdiction"/>
<restriction href="#d24e116" refersTo="#extent-e+w+s" type="jurisdiction"/>
<restriction href="#d24e129" refersTo="#extent-e+w+s" type="jurisdiction"/>
<restriction refersTo="#period-from-2020-12-31" type="jurisdiction"/>
<restriction href="#schedules" refersTo="#period-from-2018-09-01" type="jurisdiction"/>
<restriction href="#schedule-1" refersTo="#period-from-2000-07-17" type="jurisdiction"/>
<restriction href="#d24e80" refersTo="#period-from-2000-07-17" type="jurisdiction"/>
<restriction href="#d24e87" refersTo="#period-from-2000-07-17" type="jurisdiction"/>
<restriction href="#d24e97" refersTo="#period-from-2000-07-17" type="jurisdiction"/>
<restriction href="#d24e116" refersTo="#period-from-2000-07-17" type="jurisdiction"/>
<restriction href="#d24e129" refersTo="#period-from-2000-07-17" type="jurisdiction"/>
</restrictions>
<otherAnalysis source="">
<uk:commentary href="#schedule-1" refersTo="#key-7a43fb0faf8c8795dd26b8a2c3910c0b"/>
</otherAnalysis>
</analysis>
<temporalData source="#">
<temporalGroup eId="period-from-2000-07-17">
<timeInterval start="#date-2000-07-17" refersTo="#"/>
</temporalGroup>
<temporalGroup eId="period-from-2018-09-01">
<timeInterval start="#date-2018-09-01" refersTo="#"/>
</temporalGroup>
<temporalGroup eId="period-from-2020-12-31">
<timeInterval start="#date-2020-12-31" refersTo="#"/>
</temporalGroup>
</temporalData>
<references source="#">
<TLCOrganization eId="united-kingdom-parliament" href="" showAs="UnitedKingdomParliament"/>
<TLCEvent eId="made" href="" showAs="Made"/>
<TLCEvent eId="laid" href="" showAs="Laid"/>
<TLCEvent eId="cif" href="" showAs="ComingIntoForce"/>
<TLCLocation eId="extent-e+w+s" href="" showAs="E+W+S"/>
</references>
<notes source="#">
<note ukl:Name="Commentary" ukl:Type="I" class="commentary I" eId="key-7a43fb0faf8c8795dd26b8a2c3910c0b" marker="I1">
<p>
<ref eId="cxb0e7q94-00183" class="subref" href="http://www.legislation.gov.uk/id/uksi/2000/388/schedule/1">Sch. 1</ref>
in force at 17.7.2000, see
<ref eId="cxb0e7q94-00184" class="subref operative" href="http://www.legislation.gov.uk/id/uksi/2000/388/regulation/1/1">reg. 1(1)</ref>
</p>
</note>
</notes>
<proprietary xmlns:ukm="http://www.legislation.gov.uk/namespaces/metadata" xmlns:dc="http://purl.org/dc/elements/1.1/" xmlns:dct="http://purl.org/dc/terms/" source="#">
<dc:identifier>http://www.legislation.gov.uk/uksi/2000/388/schedule/1</dc:identifier>
<dc:title>The Weighing Equipment (Automatic Gravimetric Filling Instruments) Regulations 2000</dc:title>
<dc:subject>Weights and measures</dc:subject>
<dc:subject scheme="SIheading">WEIGHTS AND MEASURES</dc:subject>
<dc:publisher>Statute Law Database</dc:publisher>
<dc:modified>2024-04-02</dc:modified>
<dc:contributor>Expert Participation</dc:contributor>
<dct:valid>2020-12-31</dct:valid>
<dc:description>These Regulations implement, in Great Britain, International Recommendation OIML R 61 of the Organisation Internationale de Métrologie Légale relating to automatic gravimetric filling instruments (Edition 1996 (E)), to the extent that the Recommendation applies to such instruments following the grant or renewal of a certificate of approval of a pattern by the Secretary of State under section 12 of the Weights and Measures Act 1985 (the 1985 Act).</dc:description>
<ukm:SecondaryMetadata>
<ukm:DocumentClassification>
<ukm:DocumentCategory Value="secondary"/>
<ukm:DocumentMainType Value="UnitedKingdomStatutoryInstrument"/>
<ukm:DocumentStatus Value="revised"/>
<ukm:DocumentMinorType Value="regulation"/>
</ukm:DocumentClassification>
<ukm:Year Value="2000"/>
<ukm:Number Value="388"/>
<ukm:Made Date="2000-02-15"/>
<ukm:Laid Date="2000-02-17" Class="UnitedKingdomParliament"/>
<ukm:ComingIntoForce>
<ukm:DateTime Date="2000-07-17"/>
</ukm:ComingIntoForce>
<ukm:ISBN Value="0110988116"/>
</ukm:SecondaryMetadata>
<ukm:Alternatives>
<ukm:Alternative Date="2008-11-18" URI="http://www.legislation.gov.uk/uksi/2000/388/pdfs/uksi_20000388_en.pdf" Title="Print Version"/>
</ukm:Alternatives>
<ukm:Statistics>
<ukm:TotalParagraphs Value="47"/>
<ukm:BodyParagraphs Value="16"/>
<ukm:ScheduleParagraphs Value="31"/>
<ukm:AttachmentParagraphs Value="0"/>
<ukm:TotalImages Value="1"/>
</ukm:Statistics>
</proprietary>
</meta>
<body>
<hcontainer name="schedules" eId="schedules">
<hcontainer name="schedule" eId="schedule-1" uk:target="true">
<num>
SCHEDULE 1
<authorialNote class="referenceNote">
<p>regulation 5(a)</p>
</authorialNote>
</num>
<heading>(Manner of erection and installation)</heading>
<hcontainer name="crossheading" class="schGroup7" eId="d25e80">
<heading>Cleaning and testing</heading>
<content>
<p>Every filling instrument shall be so positioned as to facilitate cleaning and testing.</p>
</content>
</hcontainer>
<hcontainer name="crossheading" class="schGroup7" eId="d25e87">
<heading>
Feeding device (Extract from Clause 3.5 of Part 1 of
<abbr title="the International Recommendation OIML R 61 of the Organisation Internationale de Métrologie Légale relating to automatic gravimetric filling instruments (Edition 1996 (E))">OIML R 61</abbr>
)
</heading>
<content>
<p>The feeding device shall be designed to provide sufficient and regular flowrate(s).</p>
</content>
</hcontainer>
<hcontainer name="crossheading" class="schGroup7" eId="d25e97">
<heading>
Load receptor (Clause 3.6 of Part 1 of
<abbr title="the International Recommendation OIML R 61 of the Organisation Internationale de Métrologie Légale relating to automatic gravimetric filling instruments (Edition 1996 (E))">OIML R 61</abbr>
)
</heading>
<content>
<p>The load receptor, and feed and discharge devices as appropriate, shall be designed to ensure that residual material retained after each discharge is negligible.</p>
<p>Instruments using the subtractive weighing principle shall be designed to ensure that residual material retained at feed from the discharge gate is negligible.</p>
<p>The load receptor shall provide access and facilities so that where necessary test weights or masses up to the maximum capacity can be placed in position, in a safe and secure manner. If these facilities are not a permanent fixture of the instrument, they must be kept in the vicinity of the instrument.</p>
<p>Manual discharge of the load receptor shall not be possible during automatic operation.</p>
</content>
</hcontainer>
<hcontainer name="crossheading" class="schGroup7" eId="d25e116">
<heading>
Equilibrium mechanism (Clause 3.8 of Part 1 of
<abbr title="the International Recommendation OIML R 61 of the Organisation Internationale de Métrologie Légale relating to automatic gravimetric filling instruments (Edition 1996 (E))">OIML R 61</abbr>
)
</heading>
<content>
<p>
The equilibrium mechanism may be provided with detachable masses which shall be either weights in accordance with
<abbr class="acronym" title="Organisation Internationale de Métrologie Légale" xml:lang="fr">OIML</abbr>
requirements or purpose designed masses of any nominal value, distinguishable by shape and identified with the filling instrument.
</p>
</content>
</hcontainer>
<hcontainer name="crossheading" class="schGroup7" eId="d25e129">
<heading>
Installation (Extract from Clause 5.3.1 of Part 1 of
<abbr title="the International Recommendation OIML R 61 of the Organisation Internationale de Métrologie Légale relating to automatic gravimetric filling instruments (Edition 1996 (E))">OIML R 61</abbr>
)
</heading>
<content>
<p>The installation of a filling instrument shall be so designed that an automatic weighing operation will be the same whether for the purposes of testing or for use for a transaction.</p>