Print Options
PrintThe Whole
Instrument
PrintThe Whole
Schedule
PrintThis
Section
only
Changes over time for: Paragraph 4
Timeline of Changes
This timeline shows the different points in time where a change occurred. The dates will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. The first date in the timeline will usually be the earliest date when the provision came into force. In some cases the first date is 01/02/1991 (or for Northern Ireland legislation 01/01/2006). This date is our basedate. No versions before this date are available. For further information see the Editorial Practice Guide and Glossary under Help.
Status:
Point in time view as at 31/12/2020.
Changes to legislation:
There are currently no known outstanding effects for the The Materials and Articles in Contact with Food (England) Regulations 2012, Paragraph 4.
Changes to Legislation
Revised legislation carried on this site may not be fully up to date. At the current time any known changes or effects made by subsequent legislation have been applied to the text of the legislation you are viewing by the editorial team. Please see ‘Frequently Asked Questions’ for details regarding the timescales for which new effects are identified and recorded on this site.
This section has no associated Explanatory Memorandum
[4. Performance criteria of the instrumental method of analysisU.K.
(a)The detection limit for lead and cadmium must be equal to or lower than— 0,1 mg/litre for lead; 0,01 mg/litre for cadmium.
The detection limit is defined as the concentration of the element in the 4 % acetic acid solution, as referred to in paragraph 3(a) which gives a signal equal to twice the background noise of the instrument.
(b)The limit of quantification for lead and cadmium must be equal to or lower than— 0,2 mg/litre for lead; 0,02 mg/litre for cadmium.
(c)Recovery. The recovery of lead and cadmium added to the 4 % acetic acid solution, as referred to in point 3(a), must lie within 80-120 % of the added amount.
(d)Specificity. The instrumental method of analysis used must be free from matrix and spectral interferences.]