Commission Decision of 7 May 2002 on common technical specifications for in vitro-diagnostic medical devices (notified under document number C(2002) 1344) (Text with EEA relevance) (2002/364/EC) is up to date with all changes known to be in force on or before 31 October 2024. There are changes that may be brought into force at a future date.
Revised legislation carried on this site may not be fully up to date. Changes and effects are recorded by our editorial team in lists which can be found in the ‘Changes to Legislation’ area. Where those effects have yet to be applied to the text of the legislation by the editorial team they are also listed alongside the affected provisions when you open the content using the Table of Contents below.
Article 1.The technical specifications set out in the Annex to this...
COMMON TECHNICAL SPECIFICATIONS (CTS) FOR IN VITRO DIAGNOSTIC MEDICAL DEVICES
3. COMMON TECHNICAL SPECIFICATIONS (CTS) FOR PRODUCTS REFERRED TO IN ANNEX...
3.1. CTS for performance evaluation of reagents and reagent products for...
3.1.1. Devices which detect virus infections shall meet the requirements for...
3.1.2. Devices intended by the manufacturer for testing body fluids other...
3.1.3. Devices intended by the manufacturer for self-test, i.e. home use,...
3.1.4. All performance evaluations shall be carried out in direct comparison...
3.1.5. If discrepant test results are identified as part of an...
3.1.6. Performance evaluations shall be performed on a population equivalent to...
3.1.7. Positive specimens used in the performance evaluation shall be selected...
3.1.8. Sensitivity with true positives and sero-conversion samples shall be evaluated...
3.1.9. Performance evaluation of screening assays shall include 25 positive (if...
3.1.10. Negative specimens used in a performance evaluation shall be defined...
3.1.11. For performance evaluations for screening assays (Table 1) blood donor...
3.1.12. Devices shall have a specificity of at least 99,5 % on...
3.1.13. Devices shall be evaluated to establish the effect of potential...
3.1.14. For devices intended by the manufacturer to be used with...
3.1.15. For devices intended for use with plasma the performance evaluation...
3.1.16. As part of the required risk analysis the whole system...
3.1.17. If a new in vitro diagnostic medical device belonging to...
3.2. Additional requirements for HIV and HCV antigen and antibody combined...
3.3. Additional requirements for nucleic acid amplification techniques (NAT)
3.3.1. For target sequence amplification assays, a functionality control for each...
3.3.2. The analytical sensitivity or detection limit for NAT assays shall...
3.3.2a. Qualitative HIV NAT assays intended to be used to detect...
3.3.2b. Qualitative HIV NAT assays, other than virus typing assays, shall...
3.3.3. Genotype detection shall be demonstrated by appropriate primer or probe...
3.3.4. Results of quantitative NAT assays shall be traceable to international...
3.3.5. NAT assays may be used to detect virus in antibody...
3.3.6. For investigation of potential carry-over, at least five runs with...
3.3.7. The whole system failure rate leading to false-negative results shall...
3.4. CTS for the manufacturer’s release testing of reagents and reagent...
3.5. CTS for performance evaluation of reagents and reagent products for...
3.5.1. All performance evaluations shall be carried out in direct comparison...
3.5.2. If discrepant test results are identified as part of an...
3.5.3. Performance evaluations shall be performed on a population equivalent to...
3.5.4. Positive specimens used in the performance evaluation shall be selected...
3.5.5. Devices shall be evaluated to establish the effect of potential...
3.5.6. For devices intended for use with plasma the performance evaluation...
3.6. CTS for the manufacturer’s release testing of reagents and reagent...
3.7. CTS for Variant Creutzfeldt-Jakob disease (vCJD) assays for blood screening...
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
This timeline shows the different versions taken from EUR-Lex before exit day and during the implementation period as well as any subsequent versions created after the implementation period as a result of changes made by UK legislation.
The dates for the EU versions are taken from the document dates on EUR-Lex and may not always coincide with when the changes came into force for the document.
For any versions created after the implementation period as a result of changes made by UK legislation the date will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. For further information see our guide to revised legislation on Understanding Legislation.
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: