46.—(1) In addition to the requirements set out in paragraphs 44 and 45, the following specific requirements shall apply in relation to automatic catchweighers.

(2) Definitions—

“weight labeller” means an automatic catchweigher that labels individual articles with the weight value;

“weight/price labeller” means an automatic catchweigher that labels individual articles with the weight value, and price information.

(3) Accuracy Classes

(a)Instruments are divided into primary categories designated by X or Y as specified by the manufacturer.

(b)These categories are divided into four accuracy classes—

(i)XI, XII, XIII and XIV and

(ii)Y(I), Y(II), Y(a) and Y(b)

which shall be specified by the manufacturer.

(c)Category X applies to instruments used to check pre-packages made up in accordance with the requirements of Council Directive 75/106/EEC of 19 December 1974 on the approximation of the laws of the member States relating to the making-up by volume of certain pre-packaged liquids(1) and of Council Directive 76/211/EEC of 20th January 1976 on the approximation of the laws of the member States relating to the making-up by weight or by volume of certain pre-packaged products(2) applicable to pre-packages.

(d)The accuracy classes are supplemented by a factor (x) that qualifies the maximum permissible standard deviation as specified in sub-paragraph (3)(b).

(e)The manufacturer shall specify the factor (x), where (x) shall be ≤ 2 and in the form 1 × 10k, 2 × 10k or 5 × 10k, where k is a negative whole number or zero.

(f)Category Y applies to all other automatic catchweighers.

(4) MPE

(a)Mean error Category X and MPE Category Y Instruments

Table 7
Net Load (m) in verification scale intervals (c)Maximum permissible mean errorMaximum permissible error
XIY(1)XIIY(II)XIIIY(a)XIVY(b)XY
0 < m ≤ 50 0000 < m ≤ 5 0000 < m ≤ 5000 < m ≤ 50± 0.5 e±1 e
50 000 < m ≤ 200 0005 000 < m ≤ 20 000500 < m ≤ 2 00050 < m ≤ 200± 1.0 e±1.5 e
200 000 < m20 000 < m ≤ 100 0002 000 < m ≤ 10 000200 < m ≤ 1 000± 1.5 e± 2 e

(b)Standard deviation—

Maximum permissible value for the standard deviation of a class X (x) instrument is the result of the multiplication of the factor (x) by the value in Table 8 below.

Table 8
Net Load (m)Maximum permissible standard deviation for class X(1)
m ≤ 50g0.48%
50 g < m ≤ 100 g0.24 g
100 g < m ≤ 200 g0.24 %
200 g < m ≤ 300 g0.48 g
300 g < m ≤ 500 g0.16 %
500g < m ≤ 1 000 g0.8 g
1 000 g < m ≤ 10 000 g0.08 %
10 000 g < m ≤ 15 000 g8 g
15 000 g < m0.053 %

For class XI and XII (x) shall be less than 1.

For class XIII (x) shall be not greater than 1.

For class XIV (x) shall be greater than 1.

(c)Verification scale interval – single interval instruments

Table 9
Accuracy classesVerification scale intervalNumber of verification scale intervals n= Max/e
MinimumMaximum
XIY(I)0.001 g ≤ e50 000
XIIY(II)

0.001 g ≤ e ≤

0.05 g

100100 000
0.1 g ≤ e5 000100 000
XIIIY(a)0.1 g ≤ e ≤ 2 g10010 000
5 g ≤ e50010 000
XIVY(b)5 g ≤ e1001 000

(d)Verification scale interval – multi-interval instruments

Table 10
Accuracy classesVerification scale intervalNumber of verification scale intervals n=Max/e

Minimum value(1)

n = Maxi/e (i+1)

Maximum

n=Maxi/ei

(1)

For i = r the corresponding column of Table 9 applies with e replaced by er.

XIY(I)0.001 g ≤ ei50 000
XIIY(II)

0.001 g ≤ ei

0.05 g

5 000100 000
0.1 g ≤ ei5 000100 000
XIIIY(a)0.1 g ≤ ei50010 000
XIVY(b)5 g ≤ ei501 000

Where:

(i)i= 1, 2, …r

(ii)i = partial weighing range

(iii)r = total number of partial ranges

(5) Measurement Range

In specifying the measurement range for class Y instruments the manufacturer shall take account that the minimum capacity shall not be less than:

(a)class Y(I): 100 e

(b)class Y(II): 20 e for 0.001 g ≤ e ≤ 0.05 g, and 50 e for 0.1 g ≤ e

(c)class Y(a): 20 e

(d)class Y(b): 10 e

(e)Scales used for grading, e.g. postal scales and garbage weighers: 5 e

(6) Dynamic setting

(a)The dynamic setting facility shall operate within a load range specified by the manufacturer.

(b)When fitted, a dynamic setting facility that compensates for the dynamic effects of the load in motion shall be inhibited from operating outside the load range, and shall be capable of being secured.

(7) Performance under influence factors and electromagnetic disturbances

(a)The MPEs due to influence factors are—

(i)For category X instruments—

(aa)for automatic operation, as specified in Tables 7 and 8;

(bb)for static weighing in non-automatic operation, as specified in Table 7.

(ii)For category Y instruments—

(aa)for each load in automatic operation, as specified in Table 7;

(bb)for static weighing in non-automatic operation, as specified in category X in Table 7.

(b)The critical change value due to a disturbance is one verification scale interval.

(c)Temperature range

(i)For class XI and Y(I) the minimum range is 5° C.

(ii)For class XII and Y(II) the minimum range is 15° C.

(1)

OJ L 42, 15.2.75, p.1. Directive as last amended by Directive 89/676/EEC (O) L 398, 30.12.89, p. 18).

(2)

OJ L 46, 21.2.76, p.1. Directive as last amended by the EEA Agreement.

Yn ôl i’r brig

Options/Help

Print Options

Close

Mae deddfwriaeth ar gael mewn fersiynau gwahanol:

Y Diweddaraf sydd Ar Gael (diwygiedig):Y fersiwn ddiweddaraf sydd ar gael o’r ddeddfwriaeth yn cynnwys newidiadau a wnaed gan ddeddfwriaeth ddilynol ac wedi eu gweithredu gan ein tîm golygyddol. Gellir gweld y newidiadau nad ydym wedi eu gweithredu i’r testun eto yn yr ardal ‘Newidiadau i Ddeddfwriaeth’.

Gwreiddiol (Fel y’i Deddfwyd neu y’i Gwnaed): Mae'r wreiddiol fersiwn y ddeddfwriaeth fel ag yr oedd pan gafodd ei deddfu neu eu gwneud. Ni wnaed unrhyw newidiadau i’r testun.

Close

Dewisiadau Agor

Dewisiadau gwahanol i agor deddfwriaeth er mwyn gweld rhagor o gynnwys ar y sgrin ar yr un pryd

Close

Memorandwm Esboniadol

Mae Memoranda Esboniadol yn nodi datganiad byr o ddiben Offeryn Statudol ac yn rhoi gwybodaeth am ei amcan polisi a goblygiadau polisi. Maent yn ceisio gwneud yr Offeryn Statudol yn hygyrch i ddarllenwyr nad oes ganddynt gymhwyster cyfreithiol, ac maent yn cyd-fynd ag unrhyw Offeryn Statudol neu Offeryn Statudol Drafft a gyflwynwyd ger bron y Senedd o Fehefin 2004 ymlaen.

Close

Rhagor o Adnoddau

Gallwch wneud defnydd o ddogfennau atodol hanfodol a gwybodaeth ar gyfer yr eitem ddeddfwriaeth o’r tab hwn. Yn ddibynnol ar yr eitem ddeddfwriaeth sydd i’w gweld, gallai hyn gynnwys:

  • y PDF print gwreiddiol y fel deddfwyd fersiwn a ddefnyddiwyd am y copi print
  • rhestr o newidiadau a wnaed gan a/neu yn effeithio ar yr eitem hon o ddeddfwriaeth
  • manylion rhoi grym a newid cyffredinol
  • pob fformat o’r holl ddogfennau cysylltiedig
  • slipiau cywiro
  • dolenni i ddeddfwriaeth gysylltiedig ac adnoddau gwybodaeth eraill
Close

Rhagor o Adnoddau

Defnyddiwch y ddewislen hon i agor dogfennau hanfodol sy’n cyd-fynd â’r ddeddfwriaeth a gwybodaeth am yr eitem hon o ddeddfwriaeth. Gan ddibynnu ar yr eitem o ddeddfwriaeth sy’n cael ei gweld gall hyn gynnwys:

  • y PDF print gwreiddiol y fel gwnaed fersiwn a ddefnyddiwyd am y copi print
  • slipiau cywiro

liciwch ‘Gweld Mwy’ neu ddewis ‘Rhagor o Adnoddau’ am wybodaeth ychwanegol gan gynnwys

  • rhestr o newidiadau a wnaed gan a/neu yn effeithio ar yr eitem hon o ddeddfwriaeth
  • manylion rhoi grym a newid cyffredinol
  • pob fformat o’r holl ddogfennau cysylltiedig
  • dolenni i ddeddfwriaeth gysylltiedig ac adnoddau gwybodaeth eraill