The Non-automatic Weighing Instruments Regulations 2016

This adran has no associated Memorandwm Esboniadol

[F14.1.  On implementation of the procedures laid down in regulation 36, the error of indication shall not exceed the maximum permissible error of indication as shown in Table 3. In the case of digital indication the error of indication shall be corrected for the rounding error.E+W+S

The maximum permissible errors apply to the net value and tare value for all possible loads, excluding preset tare values.

Table 3
Maximum permissible errors
LoadMaximum permissible error
Class IClass IIClass IIIClass IIII
0 ≤ m ≤ 50 000 e0 ≤ m ≤ 5 000 e0 ≤ m ≤ 500 e0 ≤ m ≤ 50 e± 0,5 e
50 000 e < m ≤ 200 000 e5 000 e < m ≤ 20 000 e

500 e < m ≤

2 000 e

50 e < m ≤ 200 e± 1,0 e
200 000 e < m20 000 e < m ≤ 100 000 e2 000 e < m ≤ 10 000 e

200 e < m ≤

1 000 e

± 1,5 e]