Revised legislation carried on this site may not be fully up to date. At the current time any known changes or effects made by subsequent legislation have been applied to the text of the legislation you are viewing by the editorial team. Please see ‘Frequently Asked Questions’ for details regarding the timescales for which new effects are identified and recorded on this site.

Regulations 2 and 39(1)

[F1SCHEDULE 1FE+W+SMEASURING SYSTEMS FOR THE CONTINUOUS AND DYNAMIC MEASUREMENT OF QUANTITIES OF LIQUIDS OTHER THAN WATER (MI-005) (Annex VII to the Directive)

This Atodlen has no associated Memorandwm Esboniadol

The relevant essential requirements of Schedule 1A, the specific requirements of this Schedule and the conformity assessment procedures listed in this Schedule, apply to measuring systems intended for the continuous and dynamic measurement of quantities (volumes or masses) of liquids other than water. If appropriate, the terms ‘volume, and L’ in this Schedule can be read as: ‘mass and kg’.

DEFINITIONS

MeterAn instrument designed to measure continuously, memorise and display the quantity at metering conditions of liquid flowing through the measurement transducer in a closed, fully charged conduit.
CalculatorA part of a meter that receives the output signals from the measurement transducer(s) and possibly, from associated regulated measuring instruments and displays the measurement results.
Associated Measuring InstrumentAn instrument connected to the calculator for measuring certain quantities which are characteristic of the liquid, with a view to make a correction and/or conversion.
Conversion Device

A part of the calculator which by taking account of the characteristics of the liquid (temperature, density, etc.) measured using associated regulated measuring instruments, or stored in a memory, automatically converts:

— the volume of the liquid measured at metering conditions into a volume at base conditions and/or into mass, or

— the mass of the liquid measured at metering conditions into a volume at metering conditions and/or into a volume at base conditions

Note:

A conversion device includes the relevant associated measuring instruments.

Base conditionsThe specified conditions to which the measured quantity of liquid at metering conditions is converted.
Measuring SystemA system that comprises the meter itself and all devices required to ensure correct measurement or intended to facilitate the measuring operations.
Fuel dispenserA measuring system intended for the refuelling of motor vehicles, small boats and small aircraft.
Self-service arrangementAn arrangement that allows the customer to use a measuring system for the purpose of obtaining liquid for his own use.
Self-service deviceA specific device that is part of a self-service arrangement and which allows one of more measuring systems to perform in this self-service arrangement.
Minimum measured quantity (MMQ)The smallest quantity of liquid for which the measurement is metrologically acceptable for the measuring system.
Direct indicationThe indication, either volume or mass, corresponding to the measure and that the meter is physically capable of measuring.
Note:
The direct indication may be converted into another quantity using a conversion device.
Interruptible/non-interruptibleA measuring system is considered as interruptible/non-interruptible when the liquid flow can/cannot be stopped easily and rapidly.
Flowrate rangeThe range between the minimum flowrate (Qmin) and maximum flowrate (Qmax).

SPECIFIC REQUIREMENTS

Rated operating conditionsE+W+S

1.  The manufacturer shall specify the rated operating conditions for the instrument, in particular;

The flowrate rangeE+W+S

1.1.  The flowrate ran ge is subject to the following conditions:

(i)the flowrate range of a measuring system shall be within the flowrate range of each of its elements, in particular the meter.

(ii)meter and measuring system:

Table 1

Specific measuring systemCharacteristic of liquidMinimum ratio of Qmax: Qmin
Fuel dispensersNot Liquefied gases10: 1
Liquefied gases5: 1
Measuring systemCryogenic liquids5: 1
Measuring systems on pipeline and systems for loading shipsAll liquidsSuitable for use
All other measuring systemsAll liquids4: 1

1.2.  The properties of the liquid to be measured by the instrument by specifying the name or type of the liquid or its relevant characteristics, for example:E+W+S

  • — Temperature range;

  • — Pressure range;

  • — Density range;

  • — Viscosity range.

1.3.  The nominal value of the AC voltage supply and/or limits of the DC voltage supply.E+W+S

1.4.  The base conditions for converted values.E+W+S

This is without prejudice to the Secretary of State's obligations to require use of a temperature of 15 °C in accordance with section 12(1) of the Finance Act 1993.

Accuracy classification and maximum permissible errors (MPEs)E+W+S

2.1.  For quantities equal to or greater than 2 litres the MPE on indications is:E+W+S

Table 2

Accuracy Class
Accuracy Class
0.30.51.01.52.5
Measuring systems (A)0.3 %0.5 %1.0 %1.5 %2.5 %
Meters (B)0.2 %0.3 %0.6 %1.0 %1.5 %

2.2.  For quantities less than two litres the MPE on indications is:E+W+S

Table 3

Measured volume VMPE
V < 0.1 L4 × value in Table 2, applied to 0.1 L
0.1 L ≤ V < 0.2 L4 × value in Table 2
0.2 L ≤ V < 0.4 L2 × value in Table 2, applied to 0.4 L
0.4 L ≤ V < 1 L2 × value in Table 2
1 L ≤ V < 2 LValue in Table 2, applied to 2 L

2.3.  However, no matter what the measured quantity may be, the magnitude of the MPE is given by the greater of the following two values:E+W+S

  • — the absolute value of the MPE given in Table 2 or Table 3,

  • — the absolute value of the MPE for the minimum measured quantity (Emin).

2.4.1  For minimum measured quantities greater than or equal to 2 litres the following conditions apply:E+W+S

Condition 1

Emin shall fulfil the condition: Emin ≥ 2 R, where R is the smallest scale interval of the indication device.

Condition 2

Emin is given by the formula: Emin = (2MMQ)x (A/100) where:

  • — MMQ is the minimum measured quantity,

  • — A is the numerical value specified in line A of Table 2.

2.4.2.  For minimum measured quantities of less than two litres, the above mentioned condition 1 applies and Emin is twice the value specified in Table 3, and related to line A of Table 2.E+W+S

Converted indicationE+W+S

2.5.  In the case of a converted indication the MPEs are as in line A of Table 2.

Conversion devicesE+W+S

2.6.  MPEs on converted indications due to a conversion device are equal to ± (A — B), A and B being the values specified in Table 2.

Parts of conversion devices that can be tested separately

(a)Calculator

MPEs on quantities of liquid indications applicable to calculation, positive or negative, are equal to one-tenth of the MPEs as defined in line A of Table 2.

(b)Associated regulated measuring instruments

Associated regulated measuring instruments shall have an accuracy at least as good as the values in Table 4:

Table 4

MPE on MeasurementsAccuracy classes of the measuring system
0.30.51.01.52.5
Temperature± 0.3 °C± 0.5 °C± 1.0 °C
Pressure

Less than 1 MPa: ± 50 kPa

From 1 to 4 MPa: ± 5 %

Over 4 MPa: ± 200 kPa

Density± 1 kg/m3± 2 kg/m3± 5 kg/m3

These values apply to the indication of the characteristic quantities of the liquid displayed by the conversion device.

(c)Accuracy for calculating function

The MPE for the calculation of each characteristic quantity of the liquid, positive or negative, is equal to two fifths of the value fixed in (b).

2.7.  The requirement (a) in paragraph 2.6 applies to any calculation, not only conversion.E+W+S

2.8.  The measuring system shall not exploit the MPEs or systematically favour any party.E+W+S

Maximum permissible effect of disturbancesE+W+S

3.1.  The effect of an electromagnetic disturbance on a measuring system shall be one of the following:E+W+S

  • — the change in the measurement result is not greater than the critical change value as defined in paragraph 3.2, or

  • — the indication of the measurement result shows a momentary variation that cannot be interpreted, memorised or transmitted as a measuring result. Furthermore, in the case of an interruptible system, this can also mean the impossibility to perform any measurement, or

  • — the change in the measurement result is greater than the critical change value, in which case the measuring system shall permit the retrieval of the measuring result just before the critical change value occurred and cut off the flow.

3.2.  The critical change value is the greater of MPE/5 for a particular measured quantity or Emin.E+W+S

DurabilityE+W+S

4.  After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criterion shall be satisfied:

The variation of the measurement result after the durability test, when compared with the initial measurement result, shall not exceed the value for meters specified in line B of table 2.

SuitabilityE+W+S

5.1.  For any measured quantity relating to the same measurement, the indications provided by various devices shall not deviate one from another by more than one scale interval where devices have the same scale interval. In the case where the devices have different scale intervals, the deviation shall not be more than that of the greatest scale interval.E+W+S

However, in the case of a self-service arrangement the scale intervals of the main indicating device on the measuring system and the scale intervals of the self-service device shall be the same and results of measurement shall not deviate one from another.

5.2.  It shall not be possible to divert the measured quantity in normal conditions of use unless it is readily apparent.E+W+S

5.3.  Any percentage of air or gas not easily detectable in the liquid shall not lead to a variation of error greater than:E+W+S

  • — 0.5 % for liquids other than potable liquids and for liquids of a viscosity not exceeding 1 mPa.s, or

  • — 1 % for potable liquids and for liquids of a viscosity exceeding 1 mPa.s.

However, the allowed variation shall never be smaller than 1 % of MMQ. This value applies in the case of air or gas pockets.

Instruments for direct salesE+W+S

5.4.1.  A measuring system for direct sales shall be provided with means for resetting the display to zero.E+W+S

It shall not be possible to divert the measured quantity.

5.4.2.  The display of the quantity on which the transaction is based shall be permanent until all parties in the transaction have accepted the measurement result.E+W+S

5.4.3.  Measuring systems for direct sales shall be interruptible.E+W+S

5.4.4.  Any percentage of air or gas in the liquid shall not lead to a variation of error greater than the values specified in paragraph 5.3.E+W+S

Fuel DispensersE+W+S

5.5.1.  Displays on fuel dispensers shall not be capable of being reset to zero during a measurement.E+W+S

5.5.2.  The start of a new measurement shall be inhibited until the display has been reset to zero.E+W+S

5.5.3.  Where a measuring system is fitted with a price display, the difference between the indicated price and the price calculated from the unit price and the indicated quantity shall not exceed the price corresponding to EminHowever this difference need not be less than the smallest monetary value.E+W+S

Power supply failureE+W+S

6.  A measuring system shall either be provided with an emergency power supply device that will safeguard all measuring functions during the failure of the main power supply device or be equipped with means to save and display the data present in order to permit the conclusion of the transaction in progress and with means to stop the flow at the moment of the failure of the main power supply device.

Putting into useE+W+S

Table 5

Accuracy classTypes of Measuring system
0.3Measuring systems on pipeline
0.5All measuring systems if not differently stated elsewhere in this Table, in particular:
fuel dispensers (not for liquefied gases),
measuring systems on road tankers for liquids of low viscosity (< 20 mPa.s)
1.0Measuring systems for liquefied gases under pressure measured at a temperature equal to or above – 10 °C
Measuring systems normally in class 0.3 or 0.5 but used for liquids
whose temperature is less than – 10 °C or greater than 50 °C
whose dynamic viscosity is higher than 1,000 mPa.s
whose maximum volumetric flowrate is not higher than 20 L/h
1.5Measuring systems for liquefied gases under pressure measured at a temperature below – 10 °C (other than cryogenic liquids)
2.5Measuring systems for cryogenic liquids (temperature below – 153 °C)

Units of measurementE+W+S

8.  The metered quantity shall be displayed in millilitres, cubic centimetres, litres, cubic metres, grams, kilograms or tonnes.

CONFORMITY ASSESSMENT The conformity assessment procedures specified in the modules in Schedule 1B applicable to measuring systems for the continuous and dynamic measurement of quantities of liquids other than water that the manufacturer can choose between are:

(a)B and F;

(b)B and D;

(c)H1; or

(d)G.]

Yn ôl i’r brig

Options/Help

Print Options

You have chosen to open The Whole Instrument

The Whole Instrument you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?

You have chosen to open The Whole Instrument as a PDF

The Whole Instrument you have selected contains over 200 provisions and might take some time to download.

Would you like to continue?

You have chosen to open yr Offeryn Cyfan

Yr Offeryn Cyfan you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?

You have chosen to open Schedules only

Y Rhestrau you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?

Close

Mae deddfwriaeth ar gael mewn fersiynau gwahanol:

Y Diweddaraf sydd Ar Gael (diwygiedig):Y fersiwn ddiweddaraf sydd ar gael o’r ddeddfwriaeth yn cynnwys newidiadau a wnaed gan ddeddfwriaeth ddilynol ac wedi eu gweithredu gan ein tîm golygyddol. Gellir gweld y newidiadau nad ydym wedi eu gweithredu i’r testun eto yn yr ardal ‘Newidiadau i Ddeddfwriaeth’.

Gwreiddiol (Fel y’i Deddfwyd neu y’i Gwnaed): Mae'r wreiddiol fersiwn y ddeddfwriaeth fel ag yr oedd pan gafodd ei deddfu neu eu gwneud. Ni wnaed unrhyw newidiadau i’r testun.

Close

Gweler y wybodaeth ychwanegol ochr yn ochr â’r cynnwys

Rhychwant ddaearyddol: Indicates the geographical area that this provision applies to. For further information see ‘Frequently Asked Questions’.

Dangos Llinell Amser Newidiadau: See how this legislation has or could change over time. Turning this feature on will show extra navigation options to go to these specific points in time. Return to the latest available version by using the controls above in the What Version box.

Close

Dewisiadau Agor

Dewisiadau gwahanol i agor deddfwriaeth er mwyn gweld rhagor o gynnwys ar y sgrin ar yr un pryd

Close

Memorandwm Esboniadol

Mae Memoranda Esboniadol yn nodi datganiad byr o ddiben Offeryn Statudol ac yn rhoi gwybodaeth am ei amcan polisi a goblygiadau polisi. Maent yn ceisio gwneud yr Offeryn Statudol yn hygyrch i ddarllenwyr nad oes ganddynt gymhwyster cyfreithiol, ac maent yn cyd-fynd ag unrhyw Offeryn Statudol neu Offeryn Statudol Drafft a gyflwynwyd ger bron y Senedd o Fehefin 2004 ymlaen.

Close

Rhagor o Adnoddau

Gallwch wneud defnydd o ddogfennau atodol hanfodol a gwybodaeth ar gyfer yr eitem ddeddfwriaeth o’r tab hwn. Yn ddibynnol ar yr eitem ddeddfwriaeth sydd i’w gweld, gallai hyn gynnwys:

  • y PDF print gwreiddiol y fel deddfwyd fersiwn a ddefnyddiwyd am y copi print
  • rhestr o newidiadau a wnaed gan a/neu yn effeithio ar yr eitem hon o ddeddfwriaeth
  • manylion rhoi grym a newid cyffredinol
  • pob fformat o’r holl ddogfennau cysylltiedig
  • slipiau cywiro
  • dolenni i ddeddfwriaeth gysylltiedig ac adnoddau gwybodaeth eraill
Close

Asesiadau Effaith

Impact Assessments generally accompany all UK Government interventions of a regulatory nature that affect the private sector, civil society organisations and public services. They apply regardless of whether the regulation originates from a domestic or international source and can accompany primary (Acts etc) and secondary legislation (SIs). An Impact Assessment allows those with an interest in the policy area to understand:

  • Why the government is proposing to intervene;
  • The main options the government is considering, and which one is preferred;
  • How and to what extent new policies may impact on them; and,
  • The estimated costs and benefits of proposed measures.
Close

Llinell Amser Newidiadau

This timeline shows the different points in time where a change occurred. The dates will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. The first date in the timeline will usually be the earliest date when the provision came into force. In some cases the first date is 01/02/1991 (or for Northern Ireland legislation 01/01/2006). This date is our basedate. No versions before this date are available. For further information see the Editorial Practice Guide and Glossary under Help.

Close

Rhagor o Adnoddau

Defnyddiwch y ddewislen hon i agor dogfennau hanfodol sy’n cyd-fynd â’r ddeddfwriaeth a gwybodaeth am yr eitem hon o ddeddfwriaeth. Gan ddibynnu ar yr eitem o ddeddfwriaeth sy’n cael ei gweld gall hyn gynnwys:

  • y PDF print gwreiddiol y fel gwnaed fersiwn a ddefnyddiwyd am y copi print
  • slipiau cywiro

liciwch ‘Gweld Mwy’ neu ddewis ‘Rhagor o Adnoddau’ am wybodaeth ychwanegol gan gynnwys

  • rhestr o newidiadau a wnaed gan a/neu yn effeithio ar yr eitem hon o ddeddfwriaeth
  • manylion rhoi grym a newid cyffredinol
  • pob fformat o’r holl ddogfennau cysylltiedig
  • dolenni i ddeddfwriaeth gysylltiedig ac adnoddau gwybodaeth eraill