Council Directive 93/42/EECShow full title

Council Directive 93/42/EEC of 14 June 1993 concerning medical devices

I.DEFINITIONSU.K.

1.Definitions for the classification rulesU.K.

1.1.DurationU.K.
TransientU.K.

Normally intended for continuous use for less than 60 minutes.

Short termU.K.

Normally intended for continuous use for not more than 30 days.

Long termU.K.

Normally intended for continuous use for more than 30 days.

1.2.Invasive devicesU.K.
Invasive deviceU.K.

A device which, in whole or in part, penetrates inside the body, either through a body orifice or through the surface of the body.

Body orificeU.K.

Any natural opening in the body, as well as the external surface of the eyeball, or any permanent artificial opening, such as a stoma.

Surgically invasive deviceU.K.

An invasive device which penetrates inside the body through the surface of the body, with the aid or in the context of a surgical operation.

For the purposes of this Directive devices other than those referred to in the previous subparagraph and which produce penetration other than through an established body orifice, shall be treated as surgically invasive devices.

Implantable deviceU.K.

Any device which is intended:

  • to be totally introduced into the human body or,

  • to replace an epithelial surface or the surface of the eye,

by surgical intervention which is intended to remain in place after the procedure.

Any device intended to be partially introduced into the human body through surgical intervention and intended to remain in place after the procedure for at least 30 days is also considered an implantable device.

1.3.Reusable surgical instrumentU.K.

Instrument intended for surgical use by cutting, drilling, sawing, scratching, scraping, clamping, retracting, clipping or similar procedures, without connection to any active medical device and which can be reused after appropriate procedures have been carried out.

1.4.Active medical deviceU.K.

Any medical device operation of which depends on a source of electrical energy or any source of power other than that directly generated by the human body or gravity and which acts by converting this energy. Medical devices intended to transmit energy, substances or other elements between an active medical device and the patient, without any significant change, are not considered to be active medical devices.[F1 Stand alone software is considered to be an active medical device.]

1.5.Active therapeutical deviceU.K.

Any active medical device, whether used alone or in combination with other medical devices, to support, modify, replace or restore biological functions or structures with a view to treatment or alleviation of an illness, injury or handicap.

1.6.Active device for diagnosisU.K.

Any active medical device, whether used alone or in combination with other medical devices, to supply information for detecting, diagnosing, monitoring or treating physiological conditions, states of health, illnesses or congenital deformities.

[F21.7. Central circulatory system U.K.

For the purposes of this Directive, central circulatory system means the following vessels:

arteriae pulmonales, aorta ascendens, arcus aorta, aorta descendens to the bifurcatio aortae, arteriae coronariae, arteria carotis communis, arteria carotis externa, arteria carotis interna, arteriae cerebrales, truncus brachiocephalicus, venae cordis, venae pulmonales, vena cava superior, vena cava inferior.]

1.8.Central nervous systemU.K.

For the purposes of this Directive, ‘central nervous system’ means brain, meninges and spinal cord.