Commission Directive 2005/31/EC of 29 April 2005 amending Council Directive 84/500/EEC as regards a declaration of compliance and performance criteria of the analytical method for ceramic articles intended to come into contact with foodstuffs (Text with EEA relevance)
When the UK left the EU, legislation.gov.uk published EU legislation that had been published by the EU up to IP completion day (31 December 2020 11.00 p.m.). On legislation.gov.uk, these items of legislation are kept up-to-date with any amendments made by the UK since then.
Legislation.gov.uk publishes the UK version. EUR-Lex publishes the EU version. The EU Exit Web Archive holds a snapshot of EUR-Lex’s version from IP completion day (31 December 2020 11.00 p.m.).
This is the original version (as it was originally adopted).
The method allows the specific migration of lead and/or cadmium to be determined.
The determination of the specific migration of lead and/or cadmium is carried out by an instrumental method of analysis that fulfils the performance criteria of point 4.
All reagents must be of analytical quality, unless otherwise specified.
Where reference is made to water, it shall always mean distilled water or water of equivalent quality.
Add 40 ml of glacial acetic acid to water and make up to 1 000 ml.
Prepare stock solutions containing 1 000 mg/litre of lead and at least 500 mg/litre of cadmium respectively in a 4 % acetic acid solution, as referred to in point 3.1.
0,1 mg/litre for lead,
0,01 mg/litre for cadmium.
The detection limit is defined as the concentration of the element in the 4 % acetic acid solution, as referred to in point 3.1, which gives a signal equal to twice the background noise of the instrument.
0,2 mg/litre for lead,
0,02 mg/litre for cadmium.
The sample must be clean and free from grease or other matter likely to affect the test.
Wash the sample in a solution containing a household liquid detergent at a temperature of approximately 40 °C. Rinse the sample first in tap-water and then in distilled water or water of equivalent quality. Drain and dry so as to avoid any stain. The surface to be tested is not to be handled after it has been cleaned.
The sample thus prepared is tested under the conditions laid down in Annex I.
Before taking the test solution for determining lead and/or cadmium, homogenise the content of the sample by an appropriate method, which avoids any loss of solution or abrasion of the surface being tested.
Carry out a blank test on the reagent used for each series of determinations.
Carry out determinations for lead and/or cadmium under appropriate conditions.”
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: