Commission Directive 2005/31/EC of 29 April 2005 amending Council Directive 84/500/EEC as regards a declaration of compliance and performance criteria of the analytical method for ceramic articles intended to come into contact with foodstuffs (Text with EEA relevance)

Print Options
PrintThe Whole
Directive
PrintThe Whole
Annex
Changes over time for:


Timeline of Changes
This timeline shows the different versions taken from EUR-Lex before exit day and during the implementation period as well as any subsequent versions created after the implementation period as a result of changes made by UK legislation.
The dates for the EU versions are taken from the document dates on EUR-Lex and may not always coincide with when the changes came into force for the document.
For any versions created after the implementation period as a result of changes made by UK legislation the date will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. For further information see our guide to revised legislation on Understanding Legislation.
Status:
EU Directives are published on this site to aid cross referencing from UK legislation. Since IP completion day (31 December 2020 11.00 p.m.) no amendments have been applied to this version.
“ANNEX IIMETHODS OF ANALYSIS FOR DETERMINATION OF THE MIGRATION OF LEAD AND CADMIUM
1.Object and field of applicationU.K.
The method allows the specific migration of lead and/or cadmium to be determined.
2.PrincipleU.K.
The determination of the specific migration of lead and/or cadmium is carried out by an instrumental method of analysis that fulfils the performance criteria of point 4.
3.ReagentsU.K.
All reagents must be of analytical quality, unless otherwise specified.
Where reference is made to water, it shall always mean distilled water or water of equivalent quality.
3.1.4 % (v/v) acetic acid, in aqueous solutionU.K.
Add 40 ml of glacial acetic acid to water and make up to 1 000 ml.
3.2.Stock solutionsU.K.
Prepare stock solutions containing 1 000 mg/litre of lead and at least 500 mg/litre of cadmium respectively in a 4 % acetic acid solution, as referred to in point 3.1.
4.Performance criteria of the instrumental method of analysisU.K.
4.1.The detection limit for lead and cadmium must be equal to or lower than:U.K.
The detection limit is defined as the concentration of the element in the 4 % acetic acid solution, as referred to in point 3.1, which gives a signal equal to twice the background noise of the instrument.
4.2.The limit of quantification for lead and cadmium must be equal to or lower than:U.K.
4.3. Recovery. The recovery of lead and cadmium added to the 4 % acetic acid solution, as referred to in point 3.1, must lie within 80-120 % of the added amount.U.K.
4.4. Specificity. The instrumental method of analysis used must be free from matrix and spectral interferences.U.K.
5.MethodU.K.
5.1.Preparation of the sampleU.K.
The sample must be clean and free from grease or other matter likely to affect the test.
Wash the sample in a solution containing a household liquid detergent at a temperature of approximately 40 °C. Rinse the sample first in tap-water and then in distilled water or water of equivalent quality. Drain and dry so as to avoid any stain. The surface to be tested is not to be handled after it has been cleaned.
5.2.Determination of lead and/or cadmiumU.K.
The sample thus prepared is tested under the conditions laid down in Annex I.
Before taking the test solution for determining lead and/or cadmium, homogenise the content of the sample by an appropriate method, which avoids any loss of solution or abrasion of the surface being tested.
Carry out a blank test on the reagent used for each series of determinations.
Carry out determinations for lead and/or cadmium under appropriate conditions.”