xmlns:atom="http://www.w3.org/2005/Atom" xmlns:atom="http://www.w3.org/2005/Atom"

ANNEX IIIU.K.TEST PROCEDURE

Appendix 5CALIBRATION PROCEDURE

2.CALIBRATION OF THE CVS-SYSTEMU.K.

2.3.Calibration of the Critical Flow Venturi (CFV)U.K.

Calibration of the CFV is based upon the flow equation for a critical venturi. Gas flow is a function of inlet pressure and temperature, as shown below:

where,

Kv

=

calibration coefficient

pA

=

absolute pressure at venturi inlet, kPa

T

=

temperature at venturi inlet, K

2.3.1.Data analysisU.K.

The air flowrate (Qs) at each restriction setting (minimum eight settings) shall be calculated in standard m3/min from the flowmeter data using the manufacturer's prescribed method. The calibration coefficient shall be calculated from the calibration data for each setting as follows:

where,

Qs

=

air flow rate at standard conditions (101,3 kPa, 273 K), m3/s

T

=

temperature at the venturi inlet, K

pA

=

absolute pressure at venturi inlet, kPa

To determine the range of critical flow, Kv shall be plotted as a function of venturi inlet pressure. For critical (choked) flow, Kv will have a relatively constant value. As pressure decreases (vacuum increases), the venturi becomes unchoked and Kv decreases, which indicates that the CFV is operated outside the permissible range.

For a minimum of eight points in the region of critical flow, the average Kv and the standard deviation shall be calculated. The standard deviation shall not exceed ± 0,3 % of the average KV.