Directive 2014/32/EU of the European Parliament and of the Council of 26 February 2014 on the harmonisation of the laws of the Member States relating to the making available on the market of measuring instruments (recast) (Text with EEA relevance)
When the UK left the EU, legislation.gov.uk published EU legislation that had been published by the EU up to IP completion day (31 December 2020 11.00 p.m.). On legislation.gov.uk, these items of legislation are kept up-to-date with any amendments made by the UK since then.
Legislation.gov.uk publishes the UK version. EUR-Lex publishes the EU version. The EU Exit Web Archive holds a snapshot of EUR-Lex’s version from IP completion day (31 December 2020 11.00 p.m.).
Version Superseded: 27/01/2015
EU Directives are published on this site to aid cross referencing from UK legislation. Since IP completion day (31 December 2020 11.00 p.m.) no amendments have been applied to this version.
The relevant requirements of Annex I, the specific requirements of this Annex and the conformity assessment procedures listed in this Annex, apply to water meters intended for the measurement of volumes of clean, cold or heated water in residential, commercial and light industrial use.
Water Meter | An instrument designed to measure, memorise and display the volume at metering conditions of water passing through the measurement transducer. |
Minimum Flowrate (Q1) | The lowest flowrate at which the water meter provides indications that satisfy the requirements concerning the maximum permissible errors (MPEs.) |
Transitional Flowrate (Q2) | The transitional flowrate is the flowrate value occurring between the permanent and minimum flowrates, at which the flowrate range is divided into two zones, the ‘upper zone’ and the ‘lower zone’. Each zone has a characteristic MPE. |
Permanent Flowrate (Q3) | The highest flowrate at which the water meter operates in a satisfactory manner under normal conditions of use, i.e. under steady or intermittent flow conditions. |
Overload Flowrate (Q4) | The overload flowrate is the highest flowrate at which the meter operates in a satisfactory manner for a short period of time without deteriorating. |
The manufacturer shall specify the rated operating conditions for the instrument, in particular:
The flowrate range of the water.
The values for the flowrate range shall fulfil the following conditions:
The temperature range of the water.
The values for the temperature range shall fulfil the following conditions:
0,1 °C to at least 30 °C, or
30 °C to at least 90 °C.
The meter may be designed to operate over both ranges.
The relative pressure range of the water, the range being 0,3 bar to at least 10 bar at Q3.
For the power supply: the nominal value of the AC voltage supply and/or the limits of DC supply.
2 % for water having a temperature ≤ 30 °C,
3 % for water having a temperature > 30 °C.
The meter shall not exploit the MPE or systematically favour any party.
The meter shall not exploit the MPE or systematically favour any party.
the change in the measurement result is no greater than the critical change value as defined in point 7.1.3, or
the indication of the measurement result is such that it cannot be interpreted as a valid result, such as a momentary variation that cannot be interpreted, memorised or transmitted as a measuring result.
recover to operate within MPE, and
have all measurement functions safeguarded, and
allow recovery of all measurement data present just before the disturbance.
the volume corresponding to half of the magnitude of the MPE in the upper zone on the measured volume;
the volume corresponding to the MPE on the volume corresponding to one minute at flowrate Q3.
After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criteria shall be satisfied:
The variation of the measurement result after the durability test, when compared with the initial measurement result, shall not exceed:
3 % of the metered volume between Q1 included and Q2 excluded;
1,5 % of the metered volume between Q2 included and Q4 included.
The error of indication for the volume metered after the durability test shall not exceed:
± 6 % of the metered volume between Q1 included and Q2 excluded;
± 2,5 % of the metered volume between Q2 included and Q4 included for water meters intended to meter water with a temperature between 0,1 °C and 30 °C,
± 3,5 % of the metered volume between Q2 included and Q4 included for water meters intended to meter water with a temperature between 30 °C and 90 °C.
Water meters not designed to measure reverse flow shall either prevent reverse flow or shall withstand an accidental reverse flow without any deterioration or change in metrological properties.
The conformity assessment procedures referred to in Article 17 that the manufacturer can choose between are:
B + F or B + D or H1.
The Whole Directive you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
The Schedules you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.
Point in Time: This becomes available after navigating to view revised legislation as it stood at a certain point in time via Advanced Features > Show Timeline of Changes or via a point in time advanced search.
Geographical Extent: Indicates the geographical area that this provision applies to. For further information see ‘Frequently Asked Questions’.
Show Timeline of Changes: See how this legislation has or could change over time. Turning this feature on will show extra navigation options to go to these specific points in time. Return to the latest available version by using the controls above in the What Version box.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
This timeline shows the different versions taken from EUR-Lex before exit day and during the implementation period as well as any subsequent versions created after the implementation period as a result of changes made by UK legislation.
The dates for the EU versions are taken from the document dates on EUR-Lex and may not always coincide with when the changes came into force for the document.
For any versions created after the implementation period as a result of changes made by UK legislation the date will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. For further information see our guide to revised legislation on Understanding Legislation.
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: