X or Y
as specified by the manufacturer.
XI, XII, XIII & XIIII
and
Y(I), Y(II), Y(a) & Y(b)
which shall be specified by the manufacturer.
The manufacturer shall specify the factor (x), where (x) shall be ≤ 2 and in the form 1 × 10k, 2 × 10k or 5 × 10k, where k is a negative whole number or zero.
Category Y applies to all other automatic catchweighers.
Net Load (m) in verification scale intervals (e) | Maximum permissible mean error | Maximum permissible error | |||||||
---|---|---|---|---|---|---|---|---|---|
XI | Y(I) | XII | Y(II) | XIII | Y(a) | XIIII | Y(b) | X | Y |
0 < m ≤ 50 000 | 0 < m ≤ 5 000 | 0 < m ≤ 500 | 0 < m ≤ 50 | ± 0,5 e | ± 1 e | ||||
50 000 < m ≤ 200 000 | 5 000 < m ≤ 20 000 | 500 < m ≤ 2 000 | 50 < m ≤ 200 | ± 1,0 e | ± 1,5 e | ||||
200 000 < m | 20 000 < m ≤ 100 000 | 2 000 < m ≤ 10 000 | 200 < m ≤ 1 000 | ± 1,5 e | ± 2 e |
Maximum permissible value for the standard deviation of a class X (x) instrument is the result of the multiplication of the factor (x) by the value in Table 2 below.
Net Load (m) | Maximum permissible standard deviation for class X(1) |
---|---|
m ≤ 50 g | 0,48 % |
50 g < m ≤ 100 g | 0,24 g |
100 g < m ≤ 200 g | 0,24 % |
200 g < m ≤ 300 g | 0,48 g |
300 g < m ≤ 500 g | 0,16 % |
500 g < m ≤ 1 000 g | 0,8 g |
1 000 g < m ≤ 10 000 g | 0,08 % |
10 000 g < m ≤ 15 000 g | 8 g |
15 000 g < m | 0,053 % |
For class XI and XII (x) shall be less than 1.
For class XIII (x) shall be not greater than 1.
For class XIIII (x) shall be greater than 1.
a For i = r the corresponding column of Table 3 applies with e replaced by er. | ||||
Accuracy classes | Verification scale interval | Number of verification scale intervals | ||
---|---|---|---|---|
Minimum valuea | Maximum value | |||
XI | Y(I) | 0,001 g ≤ ei | 50 000 | — |
XII | Y(II) | 0,001 g ≤ ei ≤ 0,05 g | 5 000 | 100 000 |
0,1 g ≤ ei | 5 000 | 100 000 | ||
XIII | Y(a) | 0,1 g ≤ ei | 500 | 10 000 |
XIIII | Y(b) | 5 g ≤ ei | 50 | 1 000 |
Where:
=
1, 2, … r
=
partial weighing range
=
total number of partial ranges
In specifying the measurement range for class Y instruments the manufacturer shall take account that the minimum capacity shall not be less than:
class Y(I) | : | 100 e |
class Y(II) | : | 20 e for 0,001 g ≤ e ≤ 0,05 g, and 50 e for 0,1 g ≤ e |
class Y(a) | : | 20 e |
class Y(b) | : | 10 e |
Scales used for grading, e.g. postal scales and garbage weighers | : | 5 e |
For automatic operation; as specified in Tables 1 and 2,
For static weighing in non-automatic operation; as specified in Table 1.
For each load in automatic operation; as specified in Table 1,
For static weighing in non-automatic operation; as specified for category X in Table 1.
For class XI and Y(I) the minimum range is 5 °C,
For class XII and Y(II) the minimum range is 15 °C.