[F1ANNEX X U.K. DETERMINATION OF FATTY ACID METHYL ESTERS BY GAS CHROMATOGRAPHY

PART B U.K. ANALYSIS OF FATTY ACID METHYL ESTERS BY GAS CHROMATOGRAPHY

4. PROCEDURE U.K.

The operations described in 4.1 to 4.3 are for the use of a flame-ionisation detector.

4.1. Test conditions U.K.
4.1.1. Selection of optimum operating conditions for capillary columns U.K.

Owing to the efficiency and permeability of capillary columns, the separation of the constituents and the duration of the analysis are largely dependent on the flow-rate of the carrier gas in the column. It will therefore be necessary to optimise the operating conditions by adjusting this parameter (or simply column head loss) depending on whether the aim is to improve separation or speed up analysis.

The following conditions have proved to be suitable for the separation of FAMEs (C 4 to C 26 ). Examples of chromatograms are shown in Appendix B:

Injector temperature: 250 °C
Detector temperature: 250 °C
Oven temperature: 165 °C (8 min) to 210 °C at 2 °C/min
Carrier gas hydrogen: column head pressure, 179 kPa
Total flow: 154,0 ml/min;
Split ratio: 1:100
Injection volume: 1 μl