Textual Amendments
By using the 10 μl microsyringe (3.4), take 1 μl of hexane, draw in 0,5 μl of air and then 0,5 to 1 μl of the sample solution. Raise the plunger of the syringe further, so the needle is emptied. Push the needle through the membrane of the injector and after one to two seconds, inject rapidly, and then slowly remove the needle after around five seconds. An automatic injector can be used as well.
Carry out the recording until the TMSE of the corresponding alcoholic compounds present are completely eluted. The base line must continue to meet the requirements of the corresponding operating conditions (6.2.1 or 6.2.2).]