Regulation (EU) 2017/746 of the European Parliament and of the CouncilShow full title

Regulation (EU) 2017/746 of the European Parliament and of the Council of 5 April 2017 on in vitro diagnostic medical devices and repealing Directive 98/79/EC and Commission Decision 2010/227/EU (Text with EEA relevance)

 Help about what version

What Version

 Help about UK-EU Regulation

Legislation originating from the EU

When the UK left the EU, legislation.gov.uk published EU legislation that had been published by the EU up to IP completion day (31 December 2020 11.00 p.m.). On legislation.gov.uk, these items of legislation are kept up-to-date with any amendments made by the UK since then.

Close

This item of legislation originated from the EU

Legislation.gov.uk publishes the UK version. EUR-Lex publishes the EU version. The EU Exit Web Archive holds a snapshot of EUR-Lex’s version from IP completion day (31 December 2020 11.00 p.m.).

Status:

This is the original version as it was originally adopted in the EU.
This legislation may since have been updated - see the latest available (revised) version

ANNEX VIII CLASSIFICATION RULES

1.IMPLEMENTING RULES

1.1.Application of the classification rules shall be governed by the intended purpose of the devices.

1.2.If the device in question is intended to be used in combination with another device, the classification rules shall apply separately to each of the devices.

1.3.Accessories for an in vitro diagnostic medical device shall be classified in their own right separately from the device with which they are used.

1.4.Software, which drives a device or influences the use of a device, shall fall within the same class as the device.

If the software is independent of any other device, it shall be classified in its own right.

1.5.Calibrators intended to be used with a device shall be classified in the same class as the device.

1.6.Control materials with quantitative or qualitative assigned values intended for one specific analyte or multiple analytes shall be classified in the same class as the device.

1.7.The manufacturer shall take into consideration all classification and implementation rules in order to establish the proper classification for the device.

1.8.Where a manufacturer states multiple intended purposes for a device, and as a result the device falls into more than one class, it shall be classified in the higher class.

1.9.If several classification rules apply to the same device, the rule resulting in the higher classification shall apply.

1.10.Each of the classification rules shall apply to first line assays, confirmatory assays and supplemental assays.

2.CLASSIFICATION RULES

2.1.Rule 1

Devices intended to be used for the following purposes are classified as class D:

  • detection of the presence of, or exposure to, a transmissible agent in blood, blood components, cells, tissues or organs, or in any of their derivatives, in order to assess their suitability for transfusion, transplantation or cell administration;

  • detection of the presence of, or exposure to, a transmissible agent that causes a life-threatening disease with a high or suspected high risk of propagation;

  • determining the infectious load of a life-threatening disease where monitoring is critical in the process of patient management.

2.2.Rule 2

Devices intended to be used for blood grouping, or tissue typing to ensure the immunological compatibility of blood, blood components, cells, tissue or organs that are intended for transfusion or transplantation or cell administration, are classified as class C, except when intended to determine any of the following markers:

  • ABO system [A (ABO1), B (ABO2), AB (ABO3)];

  • Rhesus system [RH1 (D), RHW1, RH2 (C), RH3 (E), RH4 (c), RH5 (e)];

  • Kell system [Kel1 (K)];

  • Kidd system [JK1 (Jka), JK2 (Jkb)];

  • Duffy system [FY1 (Fya), FY2 (Fyb)];

in which case they are classified as class D.

2.3.Rule 3

Devices are classified as class C if they are intended:

(a)

for detecting the presence of, or exposure to, a sexually transmitted agent;

(b)

for detecting the presence in cerebrospinal fluid or blood of an infectious agent without a high or suspected high risk of propagation;

(c)

for detecting the presence of an infectious agent, if there is a significant risk that an erroneous result would cause death or severe disability to the individual, foetus or embryo being tested, or to the individual's offspring;

(d)

for pre-natal screening of women in order to determine their immune status towards transmissible agents;

(e)

for determining infective disease status or immune status, where there is a risk that an erroneous result would lead to a patient management decision resulting in a life-threatening situation for the patient or for the patient's offspring;

(f)

to be used as companion diagnostics;

(g)

to be used for disease staging, where there is a risk that an erroneous result would lead to a patient management decision resulting in a life-threatening situation for the patient or for the patient's offspring;

(h)

to be used in screening, diagnosis, or staging of cancer;

(i)

for human genetic testing;

(j)

for monitoring of levels of medicinal products, substances or biological components, when there is a risk that an erroneous result will lead to a patient management decision resulting in a life-threatening situation for the patient or for the patient's offspring;

(k)

for management of patients suffering from a life-threatening disease or condition;

(l)

for screening for congenital disorders in the embryo or foetus;

(m)

for screening for congenital disorders in new-born babies where failure to detect and treat such disorders could lead to life-threatening situations or severe disabilities.

2.4.Rule 4

(a)

Devices intended for self-testing are classified as class C, except for devices for the detection of pregnancy, for fertility testing and for determining cholesterol level, and devices for the detection of glucose, erythrocytes, leucocytes and bacteria in urine, which are classified as class B.

(b)

Devices intended for near-patient testing are classified in their own right.

2.5.Rule 5

The following devices are classified as class A:

(a)

products for general laboratory use, accessories which possess no critical characteristics, buffer solutions, washing solutions, and general culture media and histological stains, intended by the manufacturer to make them suitable for in vitro diagnostic procedures relating to a specific examination;

(b)

instruments intended by the manufacturer specifically to be used for in vitro diagnostic procedures;

(c)

specimen receptacles.

2.6.Rule 6

Devices not covered by the above-mentioned classification rules are classified as class B.

2.7.Rule 7

Devices which are controls without a quantitative or qualitative assigned value are classified as class B.

Back to top

Options/Help

Print Options

You have chosen to open the Whole Regulation

The Whole Regulation you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?

You have chosen to open Schedules only

The Schedules you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.

Would you like to continue?

Close

Legislation is available in different versions:

Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.

Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.

Close

Opening Options

Different options to open legislation in order to view more content on screen at once

Close

More Resources

Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the EU Official Journal
  • lists of changes made by and/or affecting this legislation item
  • all formats of all associated documents
  • correction slips
  • links to related legislation and further information resources
Close

More Resources

Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the print copy
  • correction slips

Click 'View More' or select 'More Resources' tab for additional information including:

  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • links to related legislation and further information resources