SCHEDULE 2METHODS OF ANALYSIS

PART II

12.DETERMINATION OF COPPER

Measurement

6.4  Set up the spectrophotometer (4.1) at a wavelength of 324.8 nm using an oxidising air-acetylene flame. Spray successively, in triplicate, the standard solution (6.3.2), the sample solution and the blank test solution (6.3.1), washing the instrument through with distilled water between each spraying. Plot the calibration curve using the mean absorbances as the ordinates and the corresponding concentrations of copper in μg/ml as the abscissae.

Determine the concentration of copper in the final sample and blank solution by reference to the calibration curve.