The Weighing Equipment (Automatic Catchweighing Instruments) Regulations (Northern Ireland) 2006

Regulation 10(3)

SCHEDULE 5N.I.(Prescribed Limits of Error)

Commencement Information

I1Sch. 5 in operation at 8.5.2006, see reg. 1

Maximum permissible errors for class Y(y) instrumentsN.I.

(Principle from Clause 2.3 of OIML R51)

The maximum permissible error for any load equal to or greater than the minimum capacity (Min) and equal to or less than the maximum capacity (Max) in automatic operation shall be as specified in Table 1. (Note that the mpe-value includes the digital rounding error of the indicating device.)

Table 1

Load (m) expressed in verification scale intervals (e)Maximum permissible error for class Y(y) instruments
Class Y(a)Class Y(b)Initial verificationIn-service
0 <m ≤ 5000 <m ≤50±1.5 e±2 e
500 <m ≤ 2,00050 <m ≤ 200±2 e±3 e
2,000 <m ≤ 10,000200 <m ≤ 1,000±2.5 e±4 e

(Principle from Clause 2.5.2 of OIML R51)

The maximum permissible error for any load equal to or greater than the minimum capacity (Min) and equal to or less than the maximum capacity (Max) for static weighing in non-automatic operation shall be as specified in Table 2.

Table 2

Load (m) expressed in verification scale intervals (e)Maximum permissible error for class Y(y) instruments
Class Y(a)Class Y(b)Initial verificationIn-service
0 <m ≤ 5000 <m ≤ 50±0.5 e±1 e
500 <m ≤ 2,00050 <m ≤ 200±1 e±2 e
2,000 <m ≤ 10,000200 <m ≤ 1,000±1.5 e±3 e