8.—(1) The sensitivity of a device approved under regulation 4 or 5 must be verified before and after its use, using a sound calibrator.
(2) Where–
(a)the sensitivity of the device has been verified after the measurement of noise in accordance with paragraph (1); and
(b)the sensitivity level of the device has changed by 0.5 decibels or more from the level indicated when the sensitivity of the device was verified before the measurement of noise,
that measurement may not be used for any of the purposes set out in these Regulations.
(3) A device approved under regulation 4 or 5 and the sound calibrator used in relation to it, must be tested in accordance with the requirements of BS 7580–
(a)no more than 2 years before it is used for any of the purposes set out in these Regulations; and
(b)at least once every 2 years following that test.
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As Enacted or Made): The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: