SCHEDULE 2ANNEX I OF COUNCIL DIRECTIVE OF 20TH JUNE 1990 ON THE HARMONISATION OF THE LAWS OF MEMBER STATES RELATING TO NON-AUTOMATIC WEIGHING INSTRUMENTS

METROLOGICAL REQUIREMENTS

Units of mass

1.  The units of mass used shall be legal units within the meaning of Directive 80/181/EEC(1) as last amended by Directive 85/1/EEC(2).

Subject to compliance with this condition, the following units are permitted:

For instruments that make use of the Imperial units of mass referred to above, the relevant essential requirements specified below shall be converted to the said Imperial units, using simple interpolation.

2.  Accuracy classes

2.1 The following accuracy classes have been defined:

I special

II high

III medium

IIII ordinary

The specifications of these classes are given in Table 1.

Table 1
Accuracy classes
ClassVerification scale interval (e)Minimum capacity (Min)Number of verification scale intervals
minimum valueminimum valuemaximum value
I0,001 g ≤ e100 e50 000
II0,001 g ≤ e ≤ 0,05 g20 e100100 000
0,1 g ≤ e500 e5 000100 000
III0,1 g ≤ e ≤ 2 g20 e10010 000
5 g ≤ e20 e50010 000
IIII5 g ≤ e10 e1001 000

The minimum capacity is reduced to 5e for instruments in classes II and III for determining a conveying tariff.

2.2. Scale intervals

2.2.1.The actual scale interval (d) and the verification scale interval (e) shall be in the form:

1 × 10k, 2 × 10k or 5 × 10k mass units,

k being any integer or zero.

2.2.2.For all instruments other than those with auxiliary indicating devices: d = e

2.2.3.For instruments with auxiliary indicating devices the following conditions apply:

e = 1 × 10k g

d < e ≤ 10 d

except for instruments of class I with d < 10−4 g, for which e = 10−3 g.

3.  Classification

3.1. Instruments with one weighing range

Instruments equipped with an auxiliary indicating device shall belong to class I or class II. For these instruments the minimum capacity lower limits for these two classes are obtained from Table 1 by replacement in column 3 of the verification scale interval (e) by the actual scale interval (d).

If d < 10−4 g, the maximum capacity of class I may be less than 50 000 e.

3.2. Instruments with multiple weighing ranges

Multiple weighing ranges are permitted, provided they are clearly indicated on the instrument. Each individual weighing range is classified according to 3.1. If the weighing ranges fall into different accuracy classes the instruments shall comply with the severest of the requirements that apply for the accuracy classes in which the weighing ranges fall.

3.3. Multi-interval instruments

3.3.1.Instruments with one weighing range may have several partial weighing ranges (multi-interval instruments).

Multi-interval instruments shall not be equipped what an auxiliary indicating device.

3.3.2.Each partial weighing range i of multi-interval instruments is defined by:

— its verification scale interval eiwith e*(i+1) > ei
— its maximum capacity Maxiwith Maxi = Max
— its minimum capacity Miniwith Mini = Max( − 1)
and Min1 = Min

where:

All capacities are capacities of net load, irrespective of the value of any tare used.

3.3.3.The partial weighing ranges are classified according to Table 2. All partial weighing ranges shall fall into the same accuracy class, this class being the instrument’s accuracy class.

Table 2
Multi-interval instruments
ClassVerification scale interval (e)Minimum capacity (Min)Number of verification scale intervals
Minimum valueMinimum value(1)Maximum value
(1)

For i = r the corresponding column of Table 1 applies, with e replaced by er.

I0,001 g ≤ ei100 ei50 000
II0,001 g ≤ ei ≤ 0,05 g20 ei5 000100 000
0,1 g ≤ ei50 ei5 000100 000
III0,1 g ≤ ei20 ei50010 000
IIII5 g ≤ ei10 ei501 000

Accuracy

4.—4.1. On implementation of the procedures laid down in Article 8, the error of indication shall not exceed the maximum permissible error of indication as shown in Table 3. In case of digital indication the error of indication shall be corrected for the rounding error.

The maximum permissible errors apply to the net and tare value for all possible loads, excluded preset tare values.

Table 3
Maximum permissible errors
LoadMaximum permissible error
Class IClass IIClass IIIClass IIII
0 ≤ m ≤ 50 000 e0 ≤ m ≤ 20 000 e0 ≤ m ≤ 500 e0 ≤ m ≤ 50 e± 0,5 e
50 000 e < m ≤ 200 000 e5 000 e < m ≤ 20 000 e500 e ≤ m ≤ 2 000 e50 e < m ≤ 200 e± 1,0 e
200 000 e <20 000 e < m ≤ 100 000 e2 000 e < m ≤ 10 000 e200 e < m ≤ 1 000 e± 1,5 e

4.2. The maximum permissible errors in service are twice the maximum permissible errors fixed in section 4.1.

5.  Weighing results of an instrument shall be repeatable, and shall be reproducible by the other indicating devices used and with other methods of balancing used.

The weighing results shall be sufficiently insensitive to changes in the position of the load on the load receptor.

6.  The instruments shall react to small variations in the load.

Influence quantities and time

7.—7.1. Instruments of classes II, III and IIII, liable to be used in a tilted position, shall be sufficiently insensitive to the degree of tilting that can exist in a normal installed condition.

7.2. The instruments shall meet the metrological requirements within the temperature range specified by the manufacturer. The value of this range shall be at least equal to:

In the absence of a manufacturer’s specification, the temperature range of −10 °C to +40 °C applies.

7.3. Instruments operated from a mains power supply shall meet the metrological requirements under conditions of power supply within the limits of normal fluctuation.

Instruments operated from battery power shall indicate whenever the voltage drops below the minimum required value and shall under those circumstances either continue to function correctly or be automatically put out of service.

7.4. Electronic instruments, except those in class I and in class II if e is less than 1 g, shall meet the metrological requirements under conditions of high relative humidity at the upper limit or their temperature range.

7.5. Loading an instrument in class II, III or IIII for a prolonged period of times shall have a negligible influence on the indication at load or on the zero indication immediately after removal of the load.

7.6. Under other conditions the instruments shall either continue to function correctly or be automatically put out of service.

(1)

OJ No. L39, 15.2.1980, p.40.

(2)

OJ No. L2, 3.1.1985, p.11. Since its amendment by Directive 85/1/EEC, Directive 80/181/EEC has been further amended by Directive 89/617/EEC.