SCHEDULE 2E+W+SANNEX I TO COUNCIL DIRECTIVE 90/384/EEC of 20TH JUNE 1990 ON THE HARMONISATION OF THE LAWS OF MEMBER STATES RELATING TO NON-AUTOMATIC WEIGHING INSTRUMENTS

Commencement Information

I1Sch. 2 in force at 1.1.2001, see reg. 1(1)

METROLOGICAL REQUIREMENTSE+W+S

AccuracyE+W+S

4.1.  On implementation of the procedures laid down in Article 8, the error of indication shall not exceed the maximum permissible error of indication as shown in Table 3. In case of digital indication the error of indication shall be corrected for the rounding error.

The maximum permissible errors apply to the net and tare value for all possible loads, excluded preset tare values.

TABLE 3
Maximum permissible errors
LoadMaximum permissible error
Class IClass IIClass IIIClass IIII
0 ≤m ≤ 50 000 e0 ≤m ≤ 5 000 e0 ≤m ≤ 500 e0 ≤m ≤ 50 e± 0,5 e
50 000 e < m ≤ 200 000 e5 000 e < m ≤ 20 000 e500 e < m ≤ 2 000 e50 e < m ≤ 200 e± 1,0 e
200 000 e < m20 000 e < m ≤ 100 000 e2 000 e < m ≤ 10 000 e200 e < m ≤ 1 000 e± 1,5 e