The Measuring Instruments (Cold-water Meters) Regulations 2006

PART IIIUSE FOR TRADE OF COLD-WATER METERS

Requirements for use for trade

14.  No person shall use for trade a cold-water meter unless—

(a)it is compliant with the essential requirements other than the provisions relating to maximum permissible errors referred to in paragraph 15 of Schedule 1;

(b)it operates within the maximum permissible errors set out in regulation 15; and

(c)the requirements of regulation 16 are complied with.

Maximum permissible error

15.—(1) Where a cold-water meter is used for trade within a flowrate range set out in column 1 of the following Table, it shall operate within the maximum permissible error specified for that flowrate range set out in column 2 of that Table.

(2) For the purposes of that Table, Q1, Q2 and Q4 have the meanings set out in paragraph 2 of Schedule 1.

Table

Column 1Column 2
Flowrate rangeMaximum permissible error as a percentage of quantity delivered
Q1 to < Q2± 6%
Q2 to and including Q4± 2.5%

Manner of use

16.—(1) Where a cold-water meter is marked with—

(a)a temperature range, it shall not be used for trade in temperatures outside that range;

(b)a flowrate range, it shall not be used for trade at a flowrate outside that range.

(2) Where a cold-water meter bears a mark which signifies the manner and purpose of use, it shall not be used for trade in a manner or for a purpose which does not accord with that marking.

(3) A cold-water meter shall not be used for trade in circumstances—

(a)in which it may be prevented from operating consistently or accurately; or

(b)which are likely prematurely to degrade its metrological characteristics.

(4) A cold-water meter shall not be used for trade unless, when adjusted, the calibration of the instrument is set as close to zero as is practicable.