Revised legislation carried on this site may not be fully up to date. At the current time any known changes or effects made by subsequent legislation have been applied to the text of the legislation you are viewing by the editorial team. Please see ‘Frequently Asked Questions’ for details regarding the timescales for which new effects are identified and recorded on this site.

Part CU.K.Measuring System for the Continuous and Dynamic Measurement of Quantities of Liquids other than Water

33.—(1) The relevant requirements of Part 1 and the specific requirements of this Part apply to measuring systems intended for the continuous and dynamic measurement of quantities (volumes or masses) of liquids other than water.U.K.

(2) If appropriate, the terms “volume” and “L” in this Part can be read as “mass” and “kg”.

DefinitionsU.K.

34.  In this Part C—

associated measuring instrument” means an instrument connected to the calculator for measuring certain quantities which are characteristic of the liquid, with a view to making a correction and/or conversion;

base conditions” means the specified conditions to which the measured quantity of liquid at metering conditions is converted;

calculator” means a part of a meter that receives the output signals from the measurement transducer(s) and possibly, from associated measuring instruments and displays the measurement results;

conversion device” means a part of the calculator which by taking account of the characteristics of the liquid (e.g. temperature, density) measured using associated measuring instruments, or stored in a memory, automatically converts—

(a)

the volume of the liquid measured at metering conditions into a volume at base conditions and/or into mass, or

(b)

the mass of the liquid measured at metering conditions into a volume at metering conditions and/or into a volume at base conditions;

Note: A conversion device includes the relevant associated measuring instrument.

direct indication” means the indication, either volume or mass, corresponding to the measure and that the meter is physically capable of measuring;

Note: The direct indication may be converted into another quantity using a conversion device.

flowrate range” means the range between the minimum flowrate (Qmin ) and maximum flowrate (Qmax);

fuel dispenser” means a measuring system intended for the refuelling of motor vehicles, small boats and small aircraft;

interruptible/non interruptible” means the liquid flow can/cannot be stopped easily and rapidly;

measuring system” means a system that comprises the meter itself and all devices required to ensure correct measurement or intended to facilitate the measuring operations;

meter” means an instrument designed to measure continuously, memorise and display the quantity at metering conditions of liquid flowing through the measurement transducer in a closed, fully charged conduit;

minimum measured quantity (MMQ)” means the smallest quantity of liquid for which the measurement is metrologically acceptable for the measuring system;

self-service arrangement” means an arrangement that allows the customer to use a measuring system for the purposes of obtaining liquid for his own use;

self-service device” means a specific device that is part of a self-service arrangement and which allows one or more measuring systems to perform in this self-service arrangement.

Rated operating conditionsU.K.

35.—(1) The manufacturer shall specify the rated operating conditions for the instrument, in particular—

(2) The flowrate range.

The flowrate range is subject to the following conditions—

(a)the flowrate range of a measuring system shall be within the flowrate range of each of its elements, in particular the meter.

(b)meter and measuring system—

Table 2
Specific measuring systemCharacteristic of liquidMinimum ratio of Qmax : Qmin
Fuel dispenserNot liquefied gases10:1
 Liquefied gases5:1
Measuring systemCryogenic liquids5:1
Measuring systems on pipeline and systems for loading shipsAll liquidsSuitable for use
All other measuring systemsAll liquids4:1

(3) The properties of the liquid to be measured by the instrument by specifying the name or type of the liquid or its relevant characteristics, for example—

(a)temperature range;

(b)pressure range;

(c)density range;

(d)viscosity range.

(4) The normal value of the AC voltage supply and/or limits of the DC voltage supply.

(5) The base conditions for converted values.

Note: Sub-paragraph (4) is without prejudice to the member States' obligations to require use of a temperature of either 15°C in accordance with Article 3(1) of Council Directive 92/81/EEC of 19th October 1992 on the harmonisation of the structures of excise duties on mineral oils M1 or, for heavy fuel oils, LPG and methane, another temperature pursuant to Article 3(2) of that Directive.

Marginal Citations

M1OJ L 316, 31.10.92, p. 12. Directive abolished by Directive 2003/96/EC (OJ L 283, 31.10.03, p.51).

Accuracy classification and maximum permissible errors (MPEs)U.K.

36.—(1) For quantities equal to or greater than 2 litres the MPE on indications is:

Table 3
 Accuracy class
0.30.51.01.52.5
Measuring systems (A)0.3%0.5%1.0%1.5%2.5%
Meters (B)0.2%0.3%0.6%1.0%1.5%

(2) For quantities less than two litres the MPE on indications is:

Table 4
Measured volume VMPE
V < 0.1 L4 × value in Table 3, applied to 0.1 L
0.1 L ≤ V < 0.2 L4 × value in Table 3
0.2 L ≤ V < 0.4 L2 × value in Table 3, applied to 0.4 L
0.4 L ≤ V < 1 L2 × value in Table 3
1 L ≤ V < 2 LValue in Table 3, applied to 2 L

(3) However, no matter what the measured quantity may be, the magnitude of the MPE is given by the greater of the following two values—

(a)the absolute value of the MPE given in Table 3 or Table 4;

(b)the absolute value of the MPE for the minimum measured quantity (Emin ).

(4)  

(a)For minimum measured quantities greater than or equal to 2 litres the following conditions apply—

(i)Condition 1—

Emin shall fulfil the condition: Emin ≥ 2 R, where R is the smallest scale interval of the indication device.

(ii)Condition 2—

Emin is given by the formula: Emin = (2MMQ) × (A/100), where—

(aa)MMQ is the minimum measured quantity,

(bb)A is the numerical value specified in line A of Table 3.

(b)For minimum measured quantities of less than two litres, condition 1 applies and Emin is twice the value specified in Table 4, and related to line A of Table 3.

(5) Converted indication—

In the case of a converted indication the MPEs are as in line A of Table 3.

(6) Conversion devices—

(a)MPEs on converted indications due to a conversion device are equal to ± (A–B), A and B being the values specified in Table 3.

(b)Parts of conversion devices that can be tested separately—

(i)Calculator

MPEs on quantities of liquid indications applicable to calculation, positive or negative, are equal to one-tenth of the MPEs as defined in line A of Table 3;

(ii)Associated measuring instruments

Associated measuring instruments shall have an accuracy at least as good as the values in Table 5—

Table 5
MPE on measurementsAccuracy classes of the measuring system
0.30.51.01.52.5
Temperature± 0.3 °C± 0.5 °C± 1.0 °C
Pressure

Less than 1 MPa: ± 50 kPa

From 1 to 4 MPa: ± 5 %

Over 4 MPa: ± 200 kPa

Density± 1kg/m3± 2kg/m3± 5kg/m3

These values apply to the indication of the characteristic quantities of the liquid displayed by the conversion device.

(iii)Accuracy for calculating function—

The MPE for the calculation of each characteristic quantity of the liquid, positive or negative, is equal to two fifths of the value fixed in (ii).

(7) The requirement (i) in sub-paragraph (6) applies to any calculation, not only conversion.

Maximum permissible effect of disturbancesU.K.

37.—(1) The effect of an electromagnetic disturbance on a measuring system shall be one of the following—

(a)the change in the measurement result is not greater than the critical change value as defined in sub-paragraph (2), or

(b)the indication of the measuring result shows a momentary variation that cannot be interpreted, memorised or transmitted as a measuring result. Furthermore, in the case of an interruptible system, this can also mean the impossibility to perform any measurement, or

(c)the change in the measurement result is greater than the critical change value, in which case the measuring system shall permit the retrieval of the measuring result just before the critical value occurred and cut off the flow.

(2) The critical change value is the greater of MPE/5 for a particular measured quantity or Emin.

DurabilityU.K.

38.  After an appropriate test, taking into account the period of time estimated by the manufacturer, has been performed, the following criterion shall be satisfied—

The variation of the measurement result after the durability test, when compared with the initial measurement result, shall not exceed the value for meters specified in line B of Table 3.

SuitabilityU.K.

39.—(1) For any measured quantity relating to the same measurement, the indications provided by various devices shall not deviate one from another by more than one scale interval where devices have the same scale interval. In the case where the devices have different scale intervals, the deviation shall not be more than that of the greatest scale interval.

(2) In the case of a self-service arrangement the scale intervals of the main indicating device on the measuring system and the scale intervals of the self-service device shall be the same and results of measurement shall not deviate one from another.

(3) It shall not be possible to divert the measured quantity in normal conditions of use unless it is readily apparent.

(4) Any percentage of air or gas not easily detectable in the liquid shall not lead to a variation of error greater than—

(a)0.5 % for liquids other than potable liquids and for liquids of a viscosity not exceeding 1 mPa.s, or

(b)1 % for potable liquids and for liquids of a viscosity exceeding 1 mPa.s.

The allowed variation shall never be smaller than 1% of MMQ. This value applies in the case of air or gas pockets.

(5) Instruments for direct sales

(a)A measuring system for direct sales shall be provided with means for resetting the display to zero.

(b)It shall not be possible to divert the measured quantity.

(c)The display of the quantity on which the transaction is based shall be permanent until all parties in the transaction have accepted the measurement result.

(d)Measuring systems for direct sales shall be interruptible.

(e)Any percentage of air or gas in the liquid shall not lead to a variation of error greater than the values specified in sub-paragraph (4).

(6) Fuel Dispensers

(a)Displays on fuel dispensers shall not be capable of being reset to zero during a measurement.

(b)The start of a new measurement shall be inhibited until the display has been reset to zero.

(c)Where a measuring system is fitted with a price display, the difference between the indicated pride and the price calculated from the unit price and the indicated quantity shall not exceed the price corresponding to Emin. This difference need not be less than the smallest monetary value.

Power supply failureU.K.

40.  A measuring system shall either be provided with an emergency power supply device that will safeguard all measuring functions during the failure of the main power supply device or be equipped with means to save and display the data present in order to permit the conclusion of the transaction in progress and with means to stop the flow at the moment of the failure of the main power supply device.

Putting into useU.K.

41.   

Table 6

1 However, Member States may require measuring systems of accuracy class 0.3 or 0.5 when used for the levying of duties on mineral oils when (un)loading ships and rail and road tankers.

Note: However, the manufacturer may specify a better accuracy for a certain type of measuring system

Accuracy ClassTypes of Measuring system
0.3Measuring systems on pipeline
0.5

All measuring systems if not differently stated elsewhere in this Table, in particular:

  • — fuel dispensers (not for liquified gases),

  • — measuring system on road tankers for liquids of low viscosity (< 20 mPa.s),

  • — measuring systems for (un)loading ships and rail and road tankers1,

  • — measuring systems for milk,

  • — measuring systems for refuelling aircraft.

1.0

Measuring system for liquefied gases under pressure measured at a temperature equal to or above -10°C

Measuring systems normally in class 0.3 or 0.5 but used for liquids—

  • — whose temperature is less than -10°C or greater than 50°C;

  • — whose dynamic viscosity is higher than 1000 mPa.s;

  • — whose maximum volumetric flowrate is not higher than 20 L/h.

1.5

Measuring systems for liquefied carbon dioxide

Measuring systems for liquefied gases under pressure measured at a temperature below –10ºC (other than cryogenic liquids)

2.5Measuring systems for cryogenic liquids (temperature below –153ºC)
Units of measurementU.K.

42.  The metered quantity shall be displayed in millilitres, cubic centimetres, litres, cubic metres, grams, kilogram or tonnes.

Back to top

Options/Help

Print Options

Close

Legislation is available in different versions:

Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.

Original (As Enacted or Made): The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.

Point in Time: This becomes available after navigating to view revised legislation as it stood at a certain point in time via Advanced Features > Show Timeline of Changes or via a point in time advanced search.

Close

See additional information alongside the content

Geographical Extent: Indicates the geographical area that this provision applies to. For further information see ‘Frequently Asked Questions’.

Show Timeline of Changes: See how this legislation has or could change over time. Turning this feature on will show extra navigation options to go to these specific points in time. Return to the latest available version by using the controls above in the What Version box.

Close

Opening Options

Different options to open legislation in order to view more content on screen at once

Close

Explanatory Memorandum

Explanatory Memorandum sets out a brief statement of the purpose of a Statutory Instrument and provides information about its policy objective and policy implications. They aim to make the Statutory Instrument accessible to readers who are not legally qualified and accompany any Statutory Instrument or Draft Statutory Instrument laid before Parliament from June 2004 onwards.

Close

More Resources

Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as enacted version that was used for the print copy
  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • correction slips
  • links to related legislation and further information resources
Close

Timeline of Changes

This timeline shows the different points in time where a change occurred. The dates will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. The first date in the timeline will usually be the earliest date when the provision came into force. In some cases the first date is 01/02/1991 (or for Northern Ireland legislation 01/01/2006). This date is our basedate. No versions before this date are available. For further information see the Editorial Practice Guide and Glossary under Help.

Close

More Resources

Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as made version that was used for the print copy
  • correction slips

Click 'View More' or select 'More Resources' tab for additional information including:

  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • links to related legislation and further information resources