Permissible effect of disturbances — Electromagnetic environment

This section has no associated Explanatory Memorandum

16.—(1) Under rated operating conditions and in the presence of a disturbance, the performance requirements shall be as set out below.

(2) Where the relevant instrument is intended to be used in a specified permanent continuous electromagnetic field the permitted performance during the radiated electromagnetic field-amplitude modulated test shall be within MPE.

(3) General

(a)As relevant instruments are directly connected to the mains supply and as mains current is also one of the measurands, a special electromagnetic environment is used for electricity meters.

(b)The relevant instrument shall comply with the electromagnetic environment E2 and the additional requirements in sub-paragraphs (4) and (5) below.

(c)The electromagnetic environment and permissible effects reflect the situation that there are disturbances of long duration which shall not affect the accuracy beyond the critical change values and transient disturbances, which may cause a temporary degradation or loss of function or performance but from which the relevant instrument shall recover and shall not affect the accuracy beyond the critical change values.

(d)When there is a foreseeable high risk due to lightning or where overhead supply networks are predominant, the metrological characteristics of the relevant instrument shall be protected.

(4) Effect of disturbances of long duration

Table 4
Critical change values for disturbances of long duration
DisturbanceCritical change values in percent for relevant instruments of class
ABC
(i)

In the case of electromechanical relevant instruments, no critical change values are defined for harmonic contents in the current circuits and for DC and harmonics in the current circuit.

Reversed phase sequence1.51.50.3
Voltage unbalance (only applicable to polyphase relevant instruments)421
Harmonic contents in the current circuit(i)10.80.5
DC and harmonics in the current circuit(i)631.5
Fast transient bursts642
Magnetic fields; HF (radiated RF) electromagnetic field; conducted disturbances introduced by radio-frequency fields; and oscillatory waves immunity321

(5) Permissible effect of transient electromagnetic phenomena

(a)The effect of an electromagnetic disturbance on a relevant instrument shall be such that during and immediately after a disturbance any output intended for testing the accuracy of the relevant instrument does not produce pulses or signals corresponding to an energy of more than the critical change value; and in reasonable time after the disturbance the relevant instrument shall—

(i)recover to operate within the MPE limits;

(ii)have all measurement functions safeguarded;

(iii)allow recovery of all measurement data present prior to the disturbance; and

(iv)not indicate a change in the registered energy of more than the critical change value.

(b)The critical change value in kWh is m · Un · Imax · 10-6 (m being the number of measuring elements of the relevant instrument, Un in Volts and Imax in Amps).

(c)For overcurrent the critical change value is 1.5 %.

Back to top

Options/Help

Print Options

Close

Legislation is available in different versions:

Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.

Original (As Enacted or Made): The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.

Close

Opening Options

Different options to open legislation in order to view more content on screen at once

Close

Explanatory Memorandum

Explanatory Memorandum sets out a brief statement of the purpose of a Statutory Instrument and provides information about its policy objective and policy implications. They aim to make the Statutory Instrument accessible to readers who are not legally qualified and accompany any Statutory Instrument or Draft Statutory Instrument laid before Parliament from June 2004 onwards.

Close

More Resources

Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as enacted version that was used for the print copy
  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • correction slips
  • links to related legislation and further information resources
Close

More Resources

Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as made version that was used for the print copy
  • correction slips

Click 'View More' or select 'More Resources' tab for additional information including:

  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • links to related legislation and further information resources