Revised legislation carried on this site may not be fully up to date. At the current time any known changes or effects made by subsequent legislation have been applied to the text of the legislation you are viewing by the editorial team. Please see ‘Frequently Asked Questions’ for details regarding the timescales for which new effects are identified and recorded on this site.
Regulation 2
Textual Amendments
F1Schs. 6-9 inserted (E.W.S.) (31.12.2020) by The Product Safety and Metrology etc. (Amendment etc.) (EU Exit) Regulations 2019 (S.I. 2019/696), reg. 1, Sch. 26 para. 45 (with Sch. 26 para. 5) (as amended by S.I. 2020/676, regs. 1(1), 2); 2020 c. 1, Sch. 5 para. 1(1)
Where an instrument includes, or is connected to, more than one indicating or printing device used for the applications listed in sub-paragraphs (a) to (f) of regulation 3(2), those devices which repeat the results of the weighing operation and which cannot influence the correct functioning of the instrument shall not be subject to the essential requirements if the weighing results are printed or recorded correctly and indelibly by a part of the instrument which meets the essential requirements and the results are accessible to both parties concerned by the measurement. However, in the case of instruments used for direct sales to the public, display and printing devices for the vendor and the customer must fulfil the essential requirements.
1. Units of massE+W+S
The units of mass used shall be the legal units within the meaning of the Weights and Measures Act 1985 relating to units of measurement F2.
Subject to compliance with this condition, the following units are permitted—
(a)SI units: kilogram, microgram, milligram, gram, tonne;
(b)imperial unit: troy ounce, if weighing precious metals;
(c)other non-SI unit: metric carat, if weighing precious stones.
For instruments that make use of the imperial unit of mass referred to above, the relevant essential requirements specified below shall be converted to that unit, using simple interpolation.
2. Accuracy classesE+W+S
2.1. The following accuracy classes have been defined—E+W+S
(a)I special
(b)II high
(c)III medium
(d)IIII ordinary
The specifications of these classes are given in Table 1.
Accuracy classes | ||||
---|---|---|---|---|
Class | Verification scale interval (e) | Minimum capacity (Min) | Number of verification scale intervals n = ((Max)/(e)) | |
minimum value | minimum value | maximum value | ||
I | 0,001 g ≤ e | 100 e | 50 000 | |
II | 0,001 g ≤ e ≤ 0,05 g | 20 e | 100 | 100 000 |
0,1 g ≤ e | 50 e | 5 000 | 100 000 | |
III | 0,1 g ≤ e ≤ 2 g | 20 e | 100 | 10 000 |
5 g ≤ e | 20 e | 500 | 10 000 | |
IIII | 5 g ≤ e | 10 e | 100 | 1 000 |
The minimum capacity is reduced to 5 e for instruments in classes II and III for determining a conveying tariff.
2.2. Scale intervalsE+W+S
2.2.1. The actual scale interval (d) and the verification scale interval (e) shall be in the form—E+W+S
1 x 10k, 2 x 10k, or 5 x 10k mass units,
k being any integer or zero.
2.2.2. For all instruments other than those with auxiliary indicating devices—E+W+S
d = e.
2.2.3. For instruments with auxiliary indicating devices the following conditions apply—E+W+S
e = 1 x 10kg;
d < e ≤ 10 d.
Those conditions do not apply for instruments of class I with d < 10–4 g, for which e = 10–3 g.
3. ClassificationE+W+S
3.1. Instruments with one weighing rangeE+W+S
Instruments equipped with an auxiliary indicating device shall belong to class I or class II. For these instruments the minimum capacity lower limits for these two classes are obtained from Table 1 by replacement in column 3 of the verification scale interval (e) by the actual scale interval (d).
If d < 10–4 g, the maximum capacity of class I may be less than 50 000 e.
3.2. Instruments with multiple weighing rangesE+W+S
Multiple weighing ranges are permitted, provided they are clearly indicated on the instrument. Each individual weighing range is classified according to point 3.1. If the weighing ranges fall into different accuracy classes the instrument shall comply with the severest of the requirements that apply for the accuracy classes in which the weighing ranges fall.
3.3. Multi-interval instrumentsE+W+S
3.3.1. Instruments with one weighing range may have several partial weighing ranges (multi-interval instruments).E+W+S
Multi-interval instruments shall not be equipped with an auxiliary indicating device.
3.3.2. Each partial weighing range i of multi-interval instruments is defined by—E+W+S
— its verification scale interval ei | with e(i + 1) > ei |
— its maximum capacity Maxi | with Maxr = Max |
— its minimum capacity Mini | with Mini = Max (i – 1) |
and Min1 = Min |
Where:
i = 1, 2, …r,
i = partial weighing range number,
r = the total number of partial weighing ranges
All capacities are capacities of net load, irrespective of the value of any tare used.
3.3.3. The partial weighing ranges are classified according to Table 2. All partial weighing ranges shall fall into the same accuracy class, that class being the instrument's accuracy class.E+W+S
Multi-level instruments | ||||
---|---|---|---|---|
i = 1, 2, …r, i = partial weighing range number, r = the total number of partial weighing ranges | ||||
Class | Verification scale interval (e) | Minimum capacity (Min) | Number of verification scale intervals | |
Minimum value | Minimum value F3 n = ((Maxi) / (e(i+1))) | Maximum value n = ((Maxi) / (ei)) | ||
I | 0,001 g ≤ ei | 100 e1 | 50 000 | |
II | 0,001 g ≤ ei ≤ 0,05 g | 20 e1 | 5 000 | 100 000 |
0,1 g ≤ ei | 50 e1 | 5 000 | 100 000 | |
III | 0,1 g ≤ ei | 20 e1 | 500 | 10 000 |
IIII | 5g ≤ ei | 10 e1 | 50 | 1 000 |
Textual Amendments
F3For I = r, the corresponding column of Table 1 applies, with e replaced by er.
4. AccuracyE+W+S
4.1. On implementation of the procedures laid down in regulation 36, the error of indication shall not exceed the maximum permissible error of indication as shown in Table 3. In the case of digital indication the error of indication shall be corrected for the rounding error.E+W+S
The maximum permissible errors apply to the net value and tare value for all possible loads, excluding preset tare values.
Maximum permissible errors | ||||
---|---|---|---|---|
Load | Maximum permissible error | |||
Class I | Class II | Class III | Class IIII | |
0 ≤ m ≤ 50 000 e | 0 ≤ m ≤ 5 000 e | 0 ≤ m ≤ 500 e | 0 ≤ m ≤ 50 e | ± 0,5 e |
50 000 e < m ≤ 200 000 e | 5 000 e < m ≤ 20 000 e | 500 e < m ≤ 2 000 e | 50 e < m ≤ 200 e | ± 1,0 e |
200 000 e < m | 20 000 e < m ≤ 100 000 e | 2 000 e < m ≤ 10 000 e | 200 e < m ≤ 1 000 e | ± 1,5 e |
4.2. The maximum permissible errors in service are twice the maximum permissible errors fixed in Section 4.1.E+W+S
5. Weighing results of an instrument shall be repeatable, and shall be reproducible by the other indicating devices used and in accordance with other methods of balancing used.E+W+S
The weighing results shall be sufficiently insensitive to changes in the position of the load on the load receptor.
6. The instrument shall react to small variations in the load.E+W+S
7. Influence quantities and timeE+W+S
7.1. Instruments of classes II, III and IIII, liable to be used in a tilted position, shall be sufficiently insensitive to the degree of tilting that can occur in normal use.E+W+S
7.2. The instruments shall meet the metrological requirements within the temperature range specified by the manufacturer. The value of this range shall be at least equal to—E+W+S
(a)5 °C for an instrument in class I;
(b)15 °C for an instrument in class II;
(c)30 °C for an instrument in class III or IIII.
In the absence of a manufacturer's specification, the temperature range of – 10 °C to + 40 °C applies.
7.3. Instruments operated from a mains power supply shall meet the metrological requirements under conditions of power supply within the limits of normal fluctuation.E+W+S
Instruments operated from battery power shall indicate whenever the voltage drops below the minimum required value and shall under those circumstances either continue to function correctly or be automatically put out of service.
7.4. Electronic instruments, except those in class I and in class II if e is less than 1 g, shall meet the metrological requirements under conditions of high relative humidity at the upper limit of their temperature range.E+W+S
7.5. Loading an instrument in class II, III or IIII for a prolonged period of time shall have a negligible influence on the indication at load or on the zero indication immediately after removal of the load.E+W+S
7.6. Under other conditions the instruments shall either continue to function correctly or be automatically put out of service.E+W+S
8. General requirementsE+W+S
8.1. Design and construction of the instruments shall be such that the instruments will preserve their metrological qualities when properly used and installed and when used in an environment for which they are intended. The value of the mass must be indicated.E+W+S
8.2. When exposed to disturbances, electronic instruments shall not display the effects of significant faults, or shall automatically detect and indicate them.E+W+S
Upon automatic detection of a significant fault, electronic instruments shall provide a visual or audible alarm that shall continue until the user takes corrective action or the fault disappears.
8.3. The requirements of points 8.1 and 8.2 shall be met on a lasting basis during a period of time that is normal in view of the intended use of such instruments.E+W+S
Digital electronic devices shall always exercise adequate control of the correct operation of the measuring process, of the indicating device, and of all data storage and data transfer.
Upon automatic detection of a significant durability error, electronic instruments shall provide a visual or audible alarm that shall continue until the user takes corrective action or the error disappears.
8.4. When external equipment is connected to an electronic instrument through an appropriate interface the metrological qualities of the instrument shall not be adversely influenced.E+W+S
8.5. The instruments shall have no characteristics likely to facilitate fraudulent use, whereas possibilities for unintentional misuse shall be minimal. Components that may not be dismantled or adjusted by the user shall be secured against such actions.E+W+S
8.6. Instruments shall be designed to permit ready execution of the statutory controls laid down by these Regulations.E+W+S
9. Indication of weighing results and other weight valuesE+W+S
The indication of the weighing results and other weight values shall be accurate, unambiguous and non-misleading and the indicating device shall permit easy reading of the indication under normal conditions of use.
The names and symbols of the units referred to in point 1 of this Schedule shall comply with the provisions of the Weights and Measures Act 1985 with the addition of the symbol for the metric carat which shall be the symbol ‘ct’.
Indication shall be impossible above the maximum capacity (Max), increased by 9 e.
An auxiliary indicating device is permitted only to the right of the decimal mark. An extended indicating device may be used only temporarily, and printing shall be inhibited during its functioning.
Secondary indications may be shown, provided that they cannot be mistaken for primary indications.
10. Printing of weighing results and other weight valuesE+W+S
Printed results shall be correct, suitably identified and unambiguous. The printing shall be clear, legible, non-erasable and durable.
11. LevellingE+W+S
When appropriate, instruments shall be fitted with a levelling device and a level indicator, sufficiently sensitive to allow proper installation.
12. ZeroingE+W+S
Instruments may be equipped with zeroing devices. The operation of these devices shall result in accurate zeroing and shall not cause incorrect measuring results.
13. Tare devices and preset tare devicesE+W+S
The instruments may have one or more tare devices and a preset tare device. The operation of the tare devices shall result in accurate zeroing and shall ensure correct net weighing. The operation of the preset tare device shall ensure correct determination of the calculated net value.
14. Instruments for direct sales to the public, with a maximum capacity not greater than 100 kg: additional requirementsE+W+S
Instruments for direct sale to the public shall show all essential information about the weighing operation and, in the case of price-indicating instruments, shall clearly show the customer the price calculation of the product to be purchased.
The price to pay, if indicated, shall be accurate.
Price-computing instruments shall display the essential indications long enough for the customer to read them properly.
Price-computing instruments may perform functions other than per-article weighing and price computation only if all indications related to all transactions are printed clearly and unambiguously and are conveniently arranged on a ticket or label for the customer.
Instruments shall bear no characteristics that can cause, directly or indirectly, indications the interpretation of which is not easy or straightforward.
Instruments shall safeguard customers against incorrect sales transactions due to their malfunctioning.
Auxiliary indicating devices and extended indicating devices are not permitted.
Supplementary devices are permitted only if they cannot lead to fraudulent use.
Instruments similar to those normally used for direct sales to the public which do not satisfy the requirements of this Section must carry near to the display the indelible marking ‘Not to be used for direct sale to the public’.
15. Price labelling instrumentsE+W+S
Price labelling instruments shall meet the requirements of price indicating instruments for direct sale to the public, as far as applicable to the instrument in question. The printing of a price label shall be impossible below a minimum capacity.]
The Whole Instrument you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
The Whole Instrument you have selected contains over 200 provisions and might take some time to download.
Would you like to continue?
The Whole Instrument you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
The Schedules you have selected contains over 200 provisions and might take some time to download. You may also experience some issues with your browser, such as an alert box that a script is taking a long time to run.
Would you like to continue?
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As Enacted or Made): The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.
Point in Time: This becomes available after navigating to view revised legislation as it stood at a certain point in time via Advanced Features > Show Timeline of Changes or via a point in time advanced search.
Geographical Extent: Indicates the geographical area that this provision applies to. For further information see ‘Frequently Asked Questions’.
Show Timeline of Changes: See how this legislation has or could change over time. Turning this feature on will show extra navigation options to go to these specific points in time. Return to the latest available version by using the controls above in the What Version box.
Explanatory Memorandum sets out a brief statement of the purpose of a Statutory Instrument and provides information about its policy objective and policy implications. They aim to make the Statutory Instrument accessible to readers who are not legally qualified and accompany any Statutory Instrument or Draft Statutory Instrument laid before Parliament from June 2004 onwards.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Impact Assessments generally accompany all UK Government interventions of a regulatory nature that affect the private sector, civil society organisations and public services. They apply regardless of whether the regulation originates from a domestic or international source and can accompany primary (Acts etc) and secondary legislation (SIs). An Impact Assessment allows those with an interest in the policy area to understand:
This timeline shows the different points in time where a change occurred. The dates will coincide with the earliest date on which the change (e.g an insertion, a repeal or a substitution) that was applied came into force. The first date in the timeline will usually be the earliest date when the provision came into force. In some cases the first date is 01/02/1991 (or for Northern Ireland legislation 01/01/2006). This date is our basedate. No versions before this date are available. For further information see the Editorial Practice Guide and Glossary under Help.
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: