F1SCHEDULE 6(Annex I to the Directive)

Annotations:

ESSENTIAL REQUIREMENTS

Metrological requirements

3

The partial weighing ranges are classified according to Table 2. All partial weighing ranges shall fall into the same accuracy class, that class being the instrument's accuracy class.

Table 2

Multi-level instruments

i  =  1, 2, …r,

i  =  partial weighing range number,

r  =  the total number of partial weighing ranges

Class

Verification scale interval (e)

Minimum capacity (Min)

Number of verification scale intervals

Minimum value

Minimum value F2

n = ((Maxi) / (e(i+1)))

Maximum value

n = ((Maxi) / (ei))

I

0,001 g ≤ ei

100 e1

50 000

II

0,001 g ≤ ei ≤ 0,05 g

20 e1

5 000

100 000

0,1 g ≤ ei

50 e1

5 000

100 000

III

0,1 g ≤ ei

20 e1

500

10 000

IIII

5g ≤ ei

10 e1

50

1 000