The Measuring Instruments Regulations 2016

[F1CHAPTER IIE+W+SAutomatic Catchweighers

1.  These categories are divided into four accuracy classes:E+W+S

Y(I), Y(II), Y(a) & Y(b)

which shall be specified by the manufacturer.

MPEE+W+S

2.1.  MPE Category Y instrumentsE+W+S

Table 1

Net Load (m) in verification scale intervals (e)Maximum permissible mean errorMaximum permissible error
Y(I)Y(II)Y(a)Y(b)StaticAutomatic
0 < m ≤ 50,0000 < m ≤ 5,0000 < m ≤ 5000 < m ≤ 50± 0.5 e± 1 e
50,000 < m ≤ 200,0005,000 < m ≤ 20,000500 < m ≤ 2,00050 < m ≤ 200± 1.0 e± 1.5 e
200,000 < m20,000 < m ≤ 100,0002,000 < m ≤ 10,000200 < m ≤ 1,000± 1.5± 2 e

Verification scale interval — single interval instrumentsE+W+S

Table 2
Accuracy classesVerification scale intervalNumber of verification scale intervalsn = Max/e
MinimumMaximum
MinimumMaximum
XIY(I)0.001 g ≤ e50,000
XIIY(II)0.001 g ≤ e ≤ 0.05 g100100,000
0.1 g ≤ e5,000100,000
XIIIY(a)0.1 g ≤ e ≤ 2 g10010,000
5 g ≤ e50010,000
XIIIIY(b)5 g ≤ e1001,000

Verification scale interval — multi-interval instrumentsE+W+S

Table 3

1 For i = r the corresponding column of Table 2 applies with e replaced by er.

Verification scale intervalNumber of verification scale intervalsn = Max/e

Minimum value1

n = Maxi /e(i+1)

For i = r the corresponding column of Table 2 applies with e replaced by er.

Maximum value

n = Maxi /ei

Y(I)0.001 g ei50,000
Y(II)0.001 g ei 0.05 g5,000100,000
0.1 g ei5,000100,000
Y(a)0.1 g ei50010,000
Y(b)5 g ei501 000

Where:

  • i =  1, 2, … r

  • i =  partial weighing range

  • r =  total number of partial ranges

Measurement RangeE+W+S

3.  In specifying the measurement range for class Y instruments the manufacturer shall take account that the minimum capacity shall not be less than:

class Y(I):100 e
class Y(II):20 e for 0.001 g ≤ e ≤ 0.05 g, and 50 e for 0.1 g ≤ e
class Y(a):20 e
class Y(b):10 e
Scales used for grading, e.g. postal scales and garbage weighers:5 e

Dynamic SettingE+W+S

4.1.  The dynamic setting facility shall operate within a load range specified by the manufacturer.E+W+S

4.2.  When fitted, a dynamic setting facility that compensates for the dynamic effects of the load in motion shall be inhibited from operating outside the load range, and shall be capable of being secured.E+W+S

Performance Under Influence Factors And Electromagnetic DisturbancesE+W+S

5.1.  The MPEs due to influence factors are:E+W+S

5.1.1.  For category Y instrumentsE+W+S

  • — For each load in automatic operation; as specified in Table 1,

  • — For static weighing in non-automatic operation; as specified in Table 1.

5.1.2.  The critical change value due to a disturbance is one verification scale interval.E+W+S

5.2.  Temperature range:E+W+S

  • — For class Y(I) the minimum range is 5 °C,

  • — For class Y(II) the minimum range is 15 °C.]