[F15.1.  The MPE due to an influence factor shall be as specified in Table 10.E+W+S

Table 10

Load (m) in verification scale intervals (d)MPE
0 < m ≤ 500± 0.5 d
500 < m ≤ 2,000± 1.0 d
2,000 < m ≤ 10,000± 1.5 d]