Search Legislation

Commission Directive 2005/50/ECShow full title

Commission Directive 2005/50/EC of 11 August 2005 on the reclassification of hip, knee and shoulder joint replacements in the framework of Council Directive 93/42/EEC concerning medical devices (Text with EEA relevance)

 Help about what version

What Version

 Help about opening options

Opening Options

 Help about UK-EU Regulation

Legislation originating from the EU

When the UK left the EU, legislation.gov.uk published EU legislation that had been published by the EU up to IP completion day (31 December 2020 11.00 p.m.). On legislation.gov.uk, these items of legislation are kept up-to-date with any amendments made by the UK since then.

Close

This item of legislation originated from the EU

Legislation.gov.uk publishes the UK version. EUR-Lex publishes the EU version. The EU Exit Web Archive holds a snapshot of EUR-Lex’s version from IP completion day (31 December 2020 11.00 p.m.).

Status:

This is the original version (as it was originally adopted).

Commission Directive 2005/50/EC

of 11 August 2005

on the reclassification of hip, knee and shoulder joint replacements in the framework of Council Directive 93/42/EEC concerning medical devices

(Text with EEA relevance)

THE COMMISSION OF THE EUROPEAN COMMUNITIES,

Having regard to the Treaty establishing the European Community,

Having regard to Council Directive 93/42/EEC of 14 June 1993 concerning medical devices(1), and in particular Article 13(1)(b) thereof,

Having regard to the request submitted by France and the United Kingdom,

Whereas:

(1) On the basis of the classification rules set out in Annex IX to Directive 93/42/EEC, total joint replacements are class IIb medical devices.

(2) France and the United Kingdom requested the classification of total joint replacements as class III medical devices by way of derogation from the provisions of Annex IX to Directive 93/42/EEC, in order to ensure an appropriate conformity assessment of total joint replacements before their placing on the market.

(3) Conformity assessment is based on a number of elements such as a proper classification, the designation and monitoring of the notified bodies and the proper implementation of the conformity assessment modules as described in Directive 93/42/EEC.

(4) Reclassification by derogation to the classification rules set out in Annex IX to Directive 93/42/EEC is indicated where the shortcomings identified due to the specific characteristics of a product will be more properly addressed under the conformity assessment procedures corresponding to the new category.

(5) Hip, knee and shoulder replacements should be distinguished from other total joint replacements, due to the particular complexity of the joint function to be restored and the consequent increased risk of failure due to the device itself.

(6) In particular, hip and knee replacements are weight-bearing and extremely sophisticated implants, for which the risk of revision surgery is significantly greater than for other joints.

(7) Shoulder implants are a more recent technique, which are subject to similar dynamic forces; their possible replacement is in principle connected with serious medical problems.

(8) Furthermore, hip, knee and shoulder replacement surgery is increasingly taking place on young people with a high life expectancy; consequently, the need for such implants to function properly over the life expectancy of the patients and to reduce revision surgery and its risks has been increased.

(9) Specific clinical data, including long term performance data are not always available for hip, knee and shoulder replacements before they are placed on the market and put into service; consequently, conclusions on clinical data collected by the manufacturer in the framework of the evaluation of the conformity of these products with the requirements concerning their characteristics and performance referred to in Sections 1 and 3 of Annex I to Directive 93/42/EEC should be subject to particular attention and examination in order to verify the appropriateness of the clinical data available.

(10) Total joint replacements can be subject to multiple modifications following their introduction into clinical use and placing on the market, as shown by hip and knee replacements on the market. However, experience has shown that what appear at first sight to be minor post-marketing changes to the design of previously trouble-free replacements can lead to serious problems due to unintended consequences, which may lead to early failure and major safety concerns.

(11) In order to achieve the optimal level of safety and health protection and to reduce the design related problems to the lowest level, the design dossier of hip, knee and shoulder replacements, including the clinical data used by the manufacturer to support the claimed performance and the subsequent post-marketing design and manufacturing changes should be examined in detail by the notified body before these devices are introduced in general clinical use.

(12) Consequently, the notified body should, under the full quality assurance system, effectively carry out an examination of the design dossier and of the changes to the approved design in accordance with point 4 of Annex II to Directive 93/42/EEC.

(13) For these reasons, it is necessary to proceed to the reclassification of hip, knee and shoulder total joint replacements as class III medical devices.

(14) It is necessary to provide for an adequate transitional period for hip, knee and shoulder total joint replacements already assessed as class IIb medical devices under the full quality assurance system of Annex II to Directive 93/42/EEC, allowing for their complementary assessment under point 4 of Annex II to the Directive.

(15) Hip, knee and shoulder total joint replacements already certified following the procedure relating to the EC type examination set out in Annex III to Directive 93/42/EEC, coupled with the procedure relating to the EC verification set out in Annex IV or the procedure relating to the EC declaration of conformity set out in Annex V to that Directive, are not affected by the present Directive as these certification schemes are the same for both class IIb and class III medical devices.

(16) It is necessary to provide for an adequate transitional period for hip, knee and shoulder total joint replacements that have already been subject to the procedure relating to the EC type examination under Annex III to Directive 93/42/EEC coupled with the procedure relating to the EC declaration of conformity set out in Annex VI to that Directive, allowing for their assessment under Annex IV or Annex V to Directive 93/42/EEC.

(17) The measures provided for in this Directive are in accordance with the opinion of the Committee on Medical Devices set up by Article 6(2) of Council Directive 90/385/EEC of 20 June 1990 on the approximation of the laws of the Member States relating to active implantable medical devices(2),

HAS ADOPTED THIS DIRECTIVE:

(1)

OJ L 169, 12.7.1993, p. 1. Directive as last amended by Regulation (EC) No 1882/2003 of the European Parliament and of the Council (OJ L 284, 31.10.2003, p. 1).

(2)

OJ L 189, 20.7.1990, p. 17. Directive as last amended by Regulation (EC) No 1882/2003 of the European Parliament and of the Council.

Back to top

Options/Help

Print Options

Close

Legislation is available in different versions:

Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.

Original (As adopted by EU): The original version of the legislation as it stood when it was first adopted in the EU. No changes have been applied to the text.

Close

Opening Options

Different options to open legislation in order to view more content on screen at once

Close

More Resources

Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the EU Official Journal
  • lists of changes made by and/or affecting this legislation item
  • all formats of all associated documents
  • correction slips
  • links to related legislation and further information resources
Close

More Resources

Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:

  • the original print PDF of the as adopted version that was used for the print copy
  • correction slips

Click 'View More' or select 'More Resources' tab for additional information including:

  • lists of changes made by and/or affecting this legislation item
  • confers power and blanket amendment details
  • all formats of all associated documents
  • links to related legislation and further information resources