Regulation 10
Regulation 10
1. Test liquid
4 % (v/v) acetic acid, in a freshly prepared aqueous solution.
2. Test conditions
(a)Carry out the test at a temperature of 22 ± 2 °C for a duration of 24 ± 0,5 hours.
(b)When the migration of lead is to be determined, cover the sample by an appropriate means of protection and expose it to the usual lighting conditions in a laboratory. When the migration of cadmium or of lead and cadmium is to be determined, cover the sample so as to ensure that the surface to be tested is kept in total darkness.
3. Filling
(a)Samples which can be filled—
Fill the article with a 4 % (v/v) acetic acid solution to a level no more than1 mm from the overflow point; the distance is measured from the upper rimof the sample. Samples with a flat or slightly sloping rim should be filled so that the distance between the surface of the liquid and the overflow point is no more than 6 mm measured along the sloping rim.
(b)Samples which cannot be filled—
The surface of the sample which is not intended to come into contact with foodstuffs is first covered with a suitable protective layer able to resist the action of the 4 % (v/v) acetic acid solution. The sample is then immersed in a recipient containing a known volume of acetic acid solution in such a way that the surface intended to come into contact with foodstuffs is completely covered by the test liquid.
4. Determination of the surface area
The surface area of the articles in Category 1 is equal to the surface area of the meniscus formed by the free liquid surface obtained by complying with the filling requirements set out in paragraph 3 above.”
Regulation 10
Regulation 10
1. Object and field of application
The method allows the specific migration of lead and/or cadmium to be determined.
2. Principle
The determination of the specific migration of lead and/or cadmium is carried out by an instrumental method of analysis that fulfils the performance criteria of paragraph 4.
3. Reagents
All reagents must be of analytical quality, unless otherwise specified.
Where reference is made to water, it means distilled water or water of equivalent quality.
(a)4 % (v/v) acetic acid, in aqueous solution.
Add 40 ml of glacial acetic acid to water and make up to 1 000 ml.
(b)Stock solutions
Prepare stock solutions containing 1 000 mg/litre of lead and at least 500 mg/litre of cadmium respectively in a 4 % acetic acid solution, as referred to in paragraph (a).
4. Performance criteria of the instrumental method of analysis
(a)The detection limit for lead and cadmium must be equal to or lower than—
0,1 mg/litre for lead;
0,01 mg/litre for cadmium.
The detection limit is defined as the concentration of the element in the 4 % acetic acid solution, as referred to in paragraph 3(a) which gives a signal equal to twice the background noise of the instrument.
(b)The limit of quantification for lead and cadmium must be equal to or lower than—
0,2 mg/litre for lead;
0,02 mg/litre for cadmium.
(c)Recovery. The recovery of lead and cadmium added to the 4 % acetic acid solution, as referred to in point 3(a), must lie within 80-120 % of the added amount.
(d)Specificity. The instrumental method of analysis used must be free from matrix and spectral interferences.
5. Method
(a)Preparation of the sample
The sample must be clean and free from grease or other matter likely to affect the test.
Wash the sample in a solution containing a household liquid detergent at a temperature of approximately 40 °C. Rinse the sample first in tap water and then in distilled water or water of equivalent quality. Drain and dry so as to avoid any stain. The surface to be tested is not to be handled after it has been cleaned.
(b)Determination of lead and/or cadmium
The sample thus prepared is tested under the conditions laid down in Schedule 2.
Before taking the test solution for determining lead and/or cadmium, homogenise the content of the sample by an appropriate method, which avoids any loss of solution or abrasion of the surface being tested.
Carry out a blank test on the reagent used for each series of determinations.
Carry out determinations for lead and/or cadmium under appropriate conditions.”
Regulation 10
Regulation 10A
1. The written declaration referred to in regulation 10A must contain the following information—
(a)the identity and address of the company which manufactures the finished ceramic article and of the importer who imports it into the United Kingdom;
(b)the identity of the ceramic article;
(c)the date of the declaration;
(d)the confirmation that the ceramic article meets relevant requirements in these Regulations and Regulation 1935/2004.
2. The written declaration must permit an easy identification of the goods for which it is issued and must be renewed when substantial changes in the production bring about changes in the migration of lead or cadmium or both.”
Latest Available (revised):The latest available updated version of the legislation incorporating changes made by subsequent legislation and applied by our editorial team. Changes we have not yet applied to the text, can be found in the ‘Changes to Legislation’ area.
Original (As Enacted or Made): The original version of the legislation as it stood when it was enacted or made. No changes have been applied to the text.
Explanatory Memorandum sets out a brief statement of the purpose of a Statutory Instrument and provides information about its policy objective and policy implications. They aim to make the Statutory Instrument accessible to readers who are not legally qualified and accompany any Statutory Instrument or Draft Statutory Instrument laid before Parliament from June 2004 onwards.
Access essential accompanying documents and information for this legislation item from this tab. Dependent on the legislation item being viewed this may include:
Use this menu to access essential accompanying documents and information for this legislation item. Dependent on the legislation item being viewed this may include:
Click 'View More' or select 'More Resources' tab for additional information including: